DOI 10.3390/healthcare11020278
Demonstration of Accuracy and Feasibility of Remotely Delivered Oximetry: A Blinded, Controlled, Real-World Study of Regional/Rural Children with Obstructive Sleep Apnoea.
Аннотация
Objectives
Evaluate diagnostic accuracy and feasibility of a mail-out home oximetry kit.
Design
Patients were referred for both the tertiary/quaternary-centre hospital-delivered oximetry (HDO) and for the mail-out remotely-delivered oximetry (RDO). Quantitative and qualitative data were collected. The COVID-19 pandemic began during this study; therefore, necessary methodological adjustments were implemented.
Setting
Patients were first evaluated in Swan Hill, Victoria. RDO kits were sent to home addresses. For the HDO, patients travelled to the Melbourne city area, received the kit, stayed overnight, and returned the kit the following morning.
Participants
All consecutive paediatric patients (aged 2−18), diagnosed by a specialist in Swan Hill with obstructive sleep apnoea (OSA) on history/examination, and booked for tonsillectomy +/− adenoidectomy, were recruited. Main outcome measures: Diagnostic accuracy (i.e., comparison of RDO to HDO results) and test delivery time (i.e., days from consent signature to oximetry delivery) were recorded. Patient travel distances for HDO collection were calculated using home/delivery address postcodes and Google® Maps data. Qualitative data were collected with two digital follow-up surveys.
Results
All 32 patients that had both the HDO and RDO had identical oximetry results. The HDO mean delivery time was 87.7 days, while the RDO mean delivery time was 23.6 days (p value: <0.001). Qualitatively, 3/28 preferred the HDO, while 25/28 preferred the RDO (n = 28).
Conclusions
The remote option is as accurate as the hospital option, strongly preferred by patients, more rapidly completed, and also an ideal investigation delivery method during certain emergencies, such as the COVID-19 pandemic.