A novel device for accurate and efficient testing for vision-threatening diabetic retinopathy.

Maa April Y, Feuer William J, Davis C Quentin, Pillow Ensa K, Brown Tara D, Caywood Rachel M, Chasan Joel E, Fransen Stephen R

СтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод пока не подготовлен

Аннотация

AIMS: To evaluate the performance of the RETeval device, a handheld instrument using flicker electroretinography (ERG) and pupillography on undilated subjects with diabetes, to detect vision-threatening diabetic retinopathy (VTDR).

METHODS

Performance was measured using a cross-sectional, single armed, non-interventional, multi-site study with Early Treatment Diabetic Retinopathy Study 7-standard field, stereo, color fundus photography as the gold standard. The 468 subjects were randomized to a calibration phase (80%), whose ERG and pupillary waveforms were used to formulate an equation correlating with the presence of VTDR, and a validation phase (20%), used to independently validate that equation. The primary outcome was the prevalence-corrected area under the receiver operating characteristic (ROC) curve for the detection of VTDR.

RESULTS

The area under the ROC curve was 0.86 for VTDR. With a sensitivity of 83%, the specificity was 78% and the negative predictive value was 99%. The average testing time was 2.3 min.

CONCLUSIONS

With a VTDR prevalence similar to that in the U.S., the RETeval device will identify about 75% of the population as not having VTDR with 99% accuracy. The device is simple to use, does not require pupil dilation, and has a short testing time.

Для цитирования: Maa April Y, Feuer William J, Davis C Quentin, Pillow Ensa K, Brown Tara D, Caywood Rachel M, Chasan Joel E, Fransen Stephen R. A novel device for accurate and efficient testing for vision-threatening diabetic retinopathy.. Journal of diabetes and its complications. 2015. DOI: 10.1016/j.jdiacomp.2015.12.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.