MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1371/journal.pone.0014330

Медицинская библиотека · PloS one · 2010

Equivalence of ELISpot assays demonstrated between major HIV network laboratories.

Gill Dilbinder K, Huang Yunda, Levine Gail L, Sambor Anna, Carter Donald K, Sato Alicia, Kopycinski Jakub, Hayes Peter, Hahn Bridget, Birungi Josephine, Tarragona-Fiol Tony, Wan Hong, Randles Mark, Cooper Andrew Raxworthy, Ssemaganda Aloysius, Clark Lorna, Kaleebu Pontiano, Self Steven G, Koup Richard, Wood Blake, McElrath M Juliana, Cox Josephine H, Hural John, Gilmour Jill

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The Comprehensive T Cell Vaccine Immune Monitoring Consortium (CTC-VIMC) was created to provide standardized immunogenicity monitoring services for HIV vaccine trials. The ex vivo interferon-gamma (IFN-γ) ELISpot is used extensively as a primary immunogenicity assay to assess T cell-based vaccine candidates in trials for infectious diseases and cancer. Two independent, GCLP-accredited central laboratories of CTC-VIMC routinely use their own standard operating procedures (SOPs) for ELISpot within two major networks of HIV vaccine trials. Studies are imperatively needed to assess the comparability of ELISpot measurements across laboratories to benefit optimal advancement of vaccine candidates.

METHODS

We describe an equivalence study of the two independently qualified IFN-g ELISpot SOPs. The study design, data collection and subsequent analysis were managed by independent statisticians to avoid subjectivity. The equivalence of both response rates and positivity calls to a given stimulus was assessed based on pre-specified acceptance criteria derived from a separate pilot study.

FINDINGS: Detection of positive responses was found to be equivalent between both laboratories. The 95% C.I. on the difference in response rates, for CMV (-1.5%, 1.5%) and CEF (-0.4%, 7.8%) responses, were both contained in the pre-specified equivalence margin of interval [-15%, 15%]. The lower bound of the 95% C.I. on the proportion of concordant positivity calls for CMV (97.2%) and CEF (89.5%) were both greater than the pre-specified margin of 70%. A third CTC-VIMC central laboratory already using one of the two SOPs also showed comparability when tested in a smaller sub-study.

INTERPRETATION: The described study procedure provides a prototypical example for the comparison of bioanalytical methods in HIV vaccine and other disease fields. This study also provides valuable and unprecedented information for future vaccine candidate evaluations on the comparison and pooling of ELISpot results generated by the CTC-VIMC central core laboratories.

Для цитирования: Gill Dilbinder K, Huang Yunda, Levine Gail L, Sambor Anna, Carter Donald K, Sato Alicia, Kopycinski Jakub, Hayes Peter, Hahn Bridget, Birungi Josephine, Tarragona-Fiol Tony, Wan Hong, Randles Mark, Cooper Andrew Raxworthy, Ssemaganda Aloysius, Clark Lorna, Kaleebu Pontiano, Self Steven G, Koup Richard, Wood Blake, McElrath M Juliana, Cox Josephine H, Hural John, Gilmour Jill. Equivalence of ELISpot assays demonstrated between major HIV network laboratories.. PloS one. 2010. DOI: 10.1371/journal.pone.0014330.