Paraprotein interference with turbidimetric gentamicin assay.

Dimeski Goce, Bassett Kendra, Brown Nigel

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Gentamicin due to its low level of resistance and rapid bactericidal activity is commonly used to treat gram-negative bacteria. However, due to its toxic effects it needs to be monitored. To date, no interference has been reported with gentamicin assays.

MATERIALS AND

METHODS

A patient with leg cellulitis and sepsis received a single dose of gentamicin and a sample was sent for gentamicin analysis. The sample showed high blank absorbance readings on Beckman DxC800 and DC800 analysers with various dilutions. A second sample was received and analysed on a Roche Cobas system to obtain a result. A third sample was received 107 hours later with the same results and this sample was then analysed neat and post ethanol precipitation on all the turbidimetric assays available on the DxC800 analyser.

RESULTS

The high blank absorbance was observed upon addition of the reactive reagents due to protein precipitation. Although not obvious from the patient protein results, it was shown the presence of high IgM paraprotein, 18.9 g/L (reference range 0.4-2.3 g/L) was the cause of precipitation, giving high blank readings. Of all the other turbidimetric assays, only vancomicin and valproate showed similar high blank absorbance readings. To be able to provide more rapid results it was shown ethanol could be used as a precipitant of proteins in both calibrators and patient samples with acceptable recovery.

CONCLUSION

IgM paraprotein was identified as the cause of interference with the gentamicin, vancomicin and valproate assays. Protein interference in these assays can be overcome by precipitation with ethanol.

Для цитирования: Dimeski Goce, Bassett Kendra, Brown Nigel. Paraprotein interference with turbidimetric gentamicin assay.. Biochemia medica. 2015. DOI: 10.11613/bm.2015.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.