DOI 10.1111/ics.12444

Медицинская библиотека · International journal of cosmetic science · 2018

Accuracy and reliability of a hand-held in vivo skin indentation device to assess skin elasticity.

Virén T, Iivarinen J T, Sarin J K, Harvima I, Mayrovitz H N

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

The aim of this study was to evaluate the performance of a hand-held indentation device for fast and reliable determination of skin stiffness.

METHODS

Device accuracy to indentation depths of 0.6 and 1.3 mm was first evaluated on plastic foam materials with mechanical properties verified by a laboratory material testing device. Subsequently, the device's sensitivity to detect age-related changes in skin stiffness was evaluated among 46 healthy women (18-79 years). Finally, the reproducibility of the method was tested with six healthy subjects.

RESULTS

High correlation was detected between indentation stiffness of reference material and Young's modulus determined with mechanical testing device (0.6 mm indenter: r = 0.97, P = 0.05; 1.3 mm indenter: r = 0.98, P = 0.04). Age-related decrease of 38% in skin stiffness was observed in healthy volunteers (P < 0.05). The coefficient of variation for 0.6 and 1.3 mm indenters was 7.4% and 8.5%, respectively. No trend related to hysteresis effect was observed from repeated measurements.

CONCLUSIONS

The presented indentation technique was accurate against the laboratory material testing device. Furthermore, skin changes related to ageing could be detected with the indentation technique. The new device was found to be feasible for monitoring skin stiffness in cosmetics and clinical conditions.

Для цитирования: Virén T, Iivarinen J T, Sarin J K, Harvima I, Mayrovitz H N. Accuracy and reliability of a hand-held in vivo skin indentation device to assess skin elasticity.. International journal of cosmetic science. 2018. DOI: 10.1111/ics.12444.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.