DOI 10.1016/j.hrthm.2023.09.015
Tailored electrocardiographic-based criteria for different pacing locations within the left bundle branch.
Аннотация
BACKGROUND
Electrocardiographic (ECG)-based criteria are used to confirm left bundle branch (LBB) pacing (LBBP), but current cutoff values have never been validated for different pacing locations.
OBJECTIVE
The purpose of this study was to describe diagnostic performance of V6-R wave peak time (RWPT), V6-V1 interpeak interval, and aVL-RWPT for different pacing sites within the LBB and to determine 100% specific values for each criterion at each pacing location.
METHODS
Consecutive patients with confirmed LBBP were selected. Population was divided into subgroups based on the site of pacing: left bundle trunk pacing (LBTP), left septal fascicular pacing (LSFP), left posterior fascicular pacing (LPFP), and left anterior fascicular pacing (LAFP).
RESULTS
A total of 147 patients with unequivocal LBB capture were analyzed. Left fascicular pacing was more frequently achieved (82.8%) than LBTP (17.2%). Diagnostic performance of V6-RWPT, V6-V1 interpeak interval, and aVL-RWPT for discrimination of LBBP was good in all subgroups. V6-RWPT cutoff values with 100% specificity (SP) for LBBP discrimination were 75 ms in LBTP, 68 ms in LPFP, 81 ms in LAFP, and 79.5 ms in LSFP. V6-V1 interpeak interval cutoff values with 100% SP for LBBP discrimination were 35.5 ms in LBTP, 53.5 ms in LPFP, 41 ms in LAFP, and 46 ms in LSFP. In LAFP, aVL-RWPT cutoff value with 100% SP for LBBP discrimination was 68 ms, but was 74 ms in LBTP, 74.5 ms in LSFP, and 73.5 ms in LPFP.
CONCLUSIONS
Tailored ECG-based criteria might be useful to confirm LBBP at different pacing locations within the LBB.