Медицинская библиотека · Advances in laboratory medicine · 2023

Evaluation of DiaSorin Liaison® calprotectin fecal assay adapted for pleural effusion.

de Paz Poves Cristina, Barneo-Caragol Clara, Cillero Sánchez Ana Isabel, Jimenez Mendiguchia Lucía, Quirós Caso Covadonga, Moreno Rodríguez María, López González Francisco J, Prieto García Mᵃ Belén

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Calprotectin (CP) is a calcium and zinc binding protein that is widely measured on faecal samples but its determination in other biological fluids might be of interest. The aim of this work was to validate the measurement of CP in pleural fluid by chemiluminescence.

METHODS

LIAISON®XL, a fully automated chemiluminescence analyzer, was used for CP quantification on pleural fluid. A validation protocol was designed using both quality control materials provided by the manufacturer and pools of pleural fluid samples. Stability, imprecision, bias, linearity, detection capability and carry over effect were evaluated.

RESULTS

CP was stable on pleural fluid at least one week, under refrigerated conditions, and four weeks at -80 °C. The observed intra- and inter-day imprecision was between 2.2 and 6.49 %, with a negative bias under 5.51 %. The linearity of the method was verified up to 2,000 ng/mL. The LoQ for the assay was 48.52 ng/mL. A statistically significant carry-over effect was observed after measuring CP concentrations above the upper limit of linearity, but given the observed magnitude, a clinically relevant impact should not be expected.

CONCLUSIONS

DiaSorin Liaison® calprotectin assay allows reliable measurement of CP in pleural fluid.

Для цитирования: de Paz Poves Cristina, Barneo-Caragol Clara, Cillero Sánchez Ana Isabel, Jimenez Mendiguchia Lucía, Quirós Caso Covadonga, Moreno Rodríguez María, López González Francisco J, Prieto García Mᵃ Belén. Evaluation of DiaSorin Liaison® calprotectin fecal assay adapted for pleural effusion.. Advances in laboratory medicine. 2023. DOI: 10.1515/almed-2023-0106.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.