MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/den.14808

Медицинская библиотека · Digestive endoscopy : official journal of the Japan Gastroenterological Endoscopy Society · 2024

Endoscopic hand suturing using a modified through-the-scope needle holder for mucosal closure after colorectal endoscopic submucosal dissection: Prospective multicenter study (with video).

Uozumi Takeshi, Abe Seiichiro, Mizuguchi Yasuhiko, Sekiguchi Masau, Toyoshima Naoya, Takamaru Hiroyuki, Yamada Masayoshi, Kobayashi Nozomu, Sadachi Ryo, Ito Sayo, Takada Kazunori, Kishida Yoshihiro, Imai Kenichiro, Hotta Kinichi, Ono Hiroyuki, Saito Yutaka

СтатьяМногоцентровое исследованиеVideo-Audio MediaПеревод пока не подготовлен

Аннотация

OBJECTIVES

Endoscopic hand suturing (EHS) is a novel technique for closing a mucosal defect after endoscopic submucosal dissection (ESD). We investigated the technical feasibility of colorectal EHS using a modified flexible through-the-scope needle holder.

METHODS

This was a prospective multicenter study conducted at two referral centers between June 2022 and April 2023. This study included colorectal neoplasms 20-50 mm in size located in the sigmoid colon or rectum. A modified flexible through-the-scope needle holder, with an increased jaw width to facilitate needle grasping, was used for colorectal EHS. The primary end-points were sustained closure rate on second-look endoscopy (SLE) performed on postoperative days 3-4 and suturing time for colorectal EHS. Secondary end-points included complete closure rate and delayed adverse events.

RESULTS

We enrolled 20 colorectal neoplasms in 20 patients, including four patients receiving antithrombotic agents. The tumor location was as follows: lower rectum (n = 8), upper rectum (n = 2), rectosigmoid colon (n = 4), and sigmoid colon (n = 6), and the median mucosal defect size was 37 mm (range, 21-65 mm). The complete closure rate was 90% (18/20 [95% confidence interval (CI) 68.3-98.8%]), and the median suturing time was 49 min (range, 23-92 min [95% CI 35-68 min]). Sustained closure rate on SLE was 85% (17/20 [95% CI 62.1-96.8%]). No delayed adverse events were observed.

CONCLUSION

EHS demonstrated a high sustained closure rate. Given the long suturing time and technical difficulty, EHS should be reserved for cases with a high risk of delayed adverse events.

Для цитирования: Uozumi Takeshi, Abe Seiichiro, Mizuguchi Yasuhiko, Sekiguchi Masau, Toyoshima Naoya, Takamaru Hiroyuki, Yamada Masayoshi, Kobayashi Nozomu, Sadachi Ryo, Ito Sayo, Takada Kazunori, Kishida Yoshihiro, Imai Kenichiro, Hotta Kinichi, Ono Hiroyuki, Saito Yutaka. Endoscopic hand suturing using a modified through-the-scope needle holder for mucosal closure after colorectal endoscopic submucosal dissection: Prospective multicenter study (with video).. Digestive endoscopy : official journal of the Japan Gastroenterological Endoscopy Society. 2024. DOI: 10.1111/den.14808.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.