International journal of laboratory hematology · 2023

Agreement between one stage and chromogenic assays in samples from patients receiving recombinant porcine FVIII (Obizur, Susoctocog-alfa).

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Guy Susan, Bowyer Annette E, Shepherd M Fiona, Maclean Rhona M, Kitchen Steve

Аннотация

INTRODUCTION: Recombinant porcine FVIII (rpFVIII) (Obizur, Susoctcog-alfa, Takeda, Japan) is licensed for the treatment of bleeding in acquired Haemophilia A (AHA). The summary of product characteristics state that monitoring should be by one stage assay (OSA) rather than chromogenic assay (CSA). CSA have been shown to underestimate activity when rpFVIII is added to plasma in vitro.

METHODS: Samples from three AHA patients (n = 21) (pre- and post rpFVIII) were assessed using FVIII:C assays; OSA methods: Actin, Actin FS, Actin FSL and Pathromtin SL performed on CS5100i (Sysmex, Kobe, Japan); APTT-SP, SynthASil and SynthAFax performed on ACL TOP (Werfen, Barcelona, Spain). CSA methods on CS5100i: Siemens Chromogenic Assay, Biophen FVIII:C, Technochrom FVIII:C; on ACL TOP: Rox Factor VIII, Coamatic Factor VIII and CRYOcheck Factor VIII.

RESULTS: OSA and CSA varied according to reagent used median OSA 61 IU/dL (range 41.5-81 IU/dL (ANOVA p < 0.0001)) median CSA 46.5 IU/dL (range of method specific medians 36.5-84 IU/dL (ANOVA p < 0.0001)). Amongst OSA, Actin FS was associated with the highest FVIII:C, APTT-SP was associated with the lowest. Variation in CSA results by different methods was also seen with highest FVIII:C levels obtained using the Technochrom FVIII:C and the lowest levels obtained with Siemens Assay.

CONCLUSION: The relationship between OSA and CSA was not consistent between method or patient. Previously there has been reports of underestimation by CSA in in vitro spiked samples. Investigation into concentration of phospholipids in the APTT reagents may explain some of these variations.