The Interday Repeatability of Parameters for the Assessment of Progressive Disease in Subjects With Less Advanced Keratoconus.

Gustafsson Ingemar, Bergström Anders, Cardiakides Anna, Ivarsen Anders, Hjortdal Jesper Østergaard

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the interday repeatability in the measurement of parameters used for the detection of progression of keratoconus by prediction limits (PL) for single measurements, and the repeatability coefficient (RC) for the mean of replicate measurements.

DESIGN

Prospective reliability analysis for cases and control eyes.

METHODS

Twenty-five eyes in 25 subjects with KC and 25 eyes in 25 healthy controls were included. Four consecutive measurements were made, 3 days apart, with a Pentacam HR tomographic instrument (denoted the Pentacam) and a Nidek ARK 560-A auto-keratometer (denoted the keratometer). Main outcome measures were the intra- and interday RC of parameters used in the detection of progression of keratoconus.

RESULTS

The most repeatable parameter obtained with the Pentacam was the curvature power of the central flat meridian (K1, 0.44 D [RC], -0.55 to 0.60 diopter [D] [PL]), followed by the central steep meridian (K2, 0.72 D [RC], -0.90 to 0.94 D [PL]). The interday repeatability of K1 and K2 was similar when using the keratometer (K1, 0.32 D [RC], -0.66 to 0.57 D [PL], K2, 0.93 D [RC], -1.36 to 1.08 D [PL]). The interday repeatability of the curvature power of the steepest point (Kmax, 0.84 D [RC], -0.90 to 1.11 D [PL]) would benefit from being stratified: RC = 0.44 D and PL = -0.49 to 0.67 D for Kmax < 49.0 D, and RC = 1.08 D and PL = -1.19 to 1.42 D for Kmax ≥ 49.0 D.

CONCLUSIONS

The interday repeatability of measurements, single or replicate, in subjects with keratoconus should be considered when diagnosing progressive disease. K1 exhibited the best intraday repeatability. Kmax benefits from being stratified according to disease severity.

Для цитирования: Gustafsson Ingemar, Bergström Anders, Cardiakides Anna, Ivarsen Anders, Hjortdal Jesper Østergaard. The Interday Repeatability of Parameters for the Assessment of Progressive Disease in Subjects With Less Advanced Keratoconus.. American journal of ophthalmology. 2021. DOI: 10.1016/j.ajo.2020.12.028.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.