Innovative dual-pump regulation in veno-arteriovenous extracorporeal membrane oxygenation: A Single-center experience.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Huang Guangyun, Liu Ying, Yuan Zhiyong, Ge Huimin, Wang Fuhua, Xing Jinyan

Аннотация

PURPOSE: To evaluate the safety and efficacy of veno-arteriovenous extracorporeal membrane oxygenation (V-AV ECMO) with a secondary centrifugal pump to control blood flow in return pathways as an alternative to the conventional Hoffman clamp.

MATERIALS AND METHODS: We conducted a retrospective review of all adult patients who underwent V-AV ECMO in our hospital from January 1, 2022, to December 31, 2023. Data, including demographic characteristics, comorbidities, and ECMO-specific information, were systematically extracted from electronic medical records.

RESULTS: Five patients, three (60 %) of whom were male, underwent V-AV ECMO. Four patients were initially placed on veno-venous extracorporeal membrane oxygenation (V-V ECMO), and one was placed on veno-arterial extracorporeal membrane oxygenation (V-A ECMO). The mean duration of V-AV ECMO support was 14.4 ± 11.4 days. Ventilator support was reduced, and fewer vasoactive drugs were administrated. Hemoglobin and platelets tended to decrease in all five patients, but D-dimer did not increase. No patients experienced active bleeding, and one patient developed oxygenator thrombosis. Two patients survived to discharge.

CONCLUSION: V-AV ECMO using a secondary centrifugal pump is feasible and likely safe without increasing the risk of ECMO-related complications such as hemolysis, thrombocytopenia, blood clot formation, or clotting factor consumption. Future prospective studies are warranted to validate these findings and optimize ECMO management protocols.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич