Comparison of dual-energy iodine and standard subtraction methods for myocardial extracellular volume quantification using cardiac computed tomography.
СтатьяСравнительное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Lorenzatti Daniel A, Filtz Annalisa, Piña Pamela, Miranda Jeirym, Ragupathi Siddarth, Contreras Yametti Felipe, Scotti Andrea, Elliott-Bailey Winsome, Cavalcante João L, Ivanc Thomas, Latib Azeem, Garcia Mario J, Slipczuk Leandro
Аннотация
BACKGROUND: Computed tomography (CT)-derived extracellular volume fraction (ECV) quantifies myocardial fibrosis noninvasively, comparable to cardiovascular magnetic resonance. The conventional subtraction method (ECV-conv) requires dedicated pre- and post-contrast acquisitions, while the dual-energy spectral method (ECV-spec) uses only the post-contrast phase. We aimed to compare these methods in patients undergoing CT planning for transcatheter aortic valve replacement (TAVR).
METHODS: We prospectively included patients undergoing CT-TAVR evaluation on a dual-energy dual-layer detector scanner. Baseline pre-contrast and equilibrium delay phase (at 5 min) prospectively ECG-triggered acquisitions were co-registered for ECV-conv calculation. Equilibrium phase spectral iodine maps were derived for ECV-spec measurement and compared with ECV-con maps using a mid-septal and a global mid-ventricular region of interest.
RESULTS: Overall, 78 patients (53 % female, mean age 77 years) were analyzed. There was a minor overestimation in global measurement by ECV-spec (31.8 ± 4.6 % vs 30.3 ± 5.3 %, p = 0.023. However, there were no significant differences between both methods for the mid-septal measurement (ECV-conv = 30.4 ± 5.3 % vs ECV-spec = 31.0 ± 5.6 %, p = 0.196).). Both methods demonstrated comparable 95 % limits of agreement, with a strong correlation for mid-septal (r = 0.75, p < 0.0001) and a moderate correlation for global measurements (r = 0.51, p < 0.0001).
CONCLUSIONS: Myocardial CT-ECV estimation using dual-energy spectral maps was comparable to the conventional subtraction method without the need of a pre-contrast acquisition. This approach may help reduce imaging and processing times, as well as minimize radiation exposure.
METHODS: We prospectively included patients undergoing CT-TAVR evaluation on a dual-energy dual-layer detector scanner. Baseline pre-contrast and equilibrium delay phase (at 5 min) prospectively ECG-triggered acquisitions were co-registered for ECV-conv calculation. Equilibrium phase spectral iodine maps were derived for ECV-spec measurement and compared with ECV-con maps using a mid-septal and a global mid-ventricular region of interest.
RESULTS: Overall, 78 patients (53 % female, mean age 77 years) were analyzed. There was a minor overestimation in global measurement by ECV-spec (31.8 ± 4.6 % vs 30.3 ± 5.3 %, p = 0.023. However, there were no significant differences between both methods for the mid-septal measurement (ECV-conv = 30.4 ± 5.3 % vs ECV-spec = 31.0 ± 5.6 %, p = 0.196).). Both methods demonstrated comparable 95 % limits of agreement, with a strong correlation for mid-septal (r = 0.75, p < 0.0001) and a moderate correlation for global measurements (r = 0.51, p < 0.0001).
CONCLUSIONS: Myocardial CT-ECV estimation using dual-energy spectral maps was comparable to the conventional subtraction method without the need of a pre-contrast acquisition. This approach may help reduce imaging and processing times, as well as minimize radiation exposure.
Предметные рубрики MeSH
HumansFemaleMaleAgedProspective StudiesContrast MediaPredictive Value of TestsReproducibility of ResultsMyocardiumFibrosisAged, 80 and overRadiographic Image Interpretation, Computer-AssistedMultidetector Computed TomographyTranscatheter Aortic Valve ReplacementCardiac-Gated Imaging TechniquesAortic Valve StenosisRadiography, Dual-Energy Scanned ProjectionAortic Valve