Battery longevity of a helix-fixation dual-chamber leadless pacemaker: results from the AVEIR DR i2i Study.

Reddy Vivek Y, Doshi Rahul, Ip James E, Defaye Pascal, Exner Derek V, Canby Robert, Shoda Morio, Bongiorni Maria G, Hindricks Gerhard, Neužil Petr, Callahan Thomas, Badie Nima, Fishler Matthew G, Knops Reinoud E

СтатьяМногоцентровое исследованиеКлиническое исследованиеПеревод пока не подготовлен

Аннотация

AIMS: A dual-chamber leadless pacemaker (LP) system that employs distinct atrial and ventricular LP devices (ALP, VLP) has been introduced to clinical practice. Proprietary, low-energy, implant-to-implant (i2i) communication at each beat enables the devices to maintain synchronous atrioventricular sensing and pacing. We evaluated device longevities and contributing factors, such as i2i communication.

METHODS AND

RESULTS

Patients meeting dual-chamber pacing indications received the dual-chamber LP system as part of a prospective, multi-centre, international clinical trial (Aveir DR i2i Study, NCT05252702). Programming and diagnostics were interrogated from all de novo, non-revised, dual-chamber programmed devices at 12 months post-implant. This analysis included 302 patients (65% male; age 70 ± 13 years; weight 80 ± 19 kg; intrinsic heart rate 55 ± 7 bpm; 58% sinus node dysfunction, 27% atrioventricular block). At 12 months, devices were programmed to dual-chamber pacing (DDD(R) or DDI(R)) at a median 60 bpm rate, median 1.25 V pulse amplitude in ALP and 1.5 V in VLP, median 0.4 ms pulse width, and median i2i signal setting level 5 out of 7. Median ALP and VLP remaining battery longevities at 12 months were 4.3 and 9.1 years, with median total ALP and VLP longevities of 5.3 and 9.9 years. Base rate, pulse amplitude, pacing percentage, event rate, impedance, and i2i setting level all exhibited significant correlations with ALP and VLP longevities (P < 0.001). Programming i2i setting levels below 7 produced the greatest longevity savings.

CONCLUSION

The first dual-chamber LP demonstrated adequate projected battery longevity after 12 months of use. Patient-specific device programming considerations, unique to leadless devices, may extend longevity.

Для цитирования: Reddy Vivek Y, Doshi Rahul, Ip James E, Defaye Pascal, Exner Derek V, Canby Robert, Shoda Morio, Bongiorni Maria G, Hindricks Gerhard, Neužil Petr, Callahan Thomas, Badie Nima, Fishler Matthew G, Knops Reinoud E. Battery longevity of a helix-fixation dual-chamber leadless pacemaker: results from the AVEIR DR i2i Study.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2025. DOI: 10.1093/europace/euaf074.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.