Comprehensive Ex Vivo Comparison of 5 Clinically Used Conduit Configurations for Valve-Sparing Aortic Root Replacement Using a 3-Dimensional-Printed Heart Simulator.

Paulsen Michael J, Imbrie-Moore Annabel M, Baiocchi Michael, Wang Hanjay, Hironaka Camille E, Lucian Haley J, Farry Justin M, Thakore Akshara D, Zhu Yuanjia, Ma Michael, MacArthur John W, Woo Y Joseph

Сравнительное исследованиеСтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПоддержка правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Many graft configurations are clinically used for valve-sparing aortic root replacement, some specifically focused on recapitulating neosinus geometry. However, the specific impact of such neosinuses on valvular and root biomechanics and the potential influence on long-term durability are unknown.

METHODS

Using a custom 3-dimenstional-printed heart simulator with porcine aortic roots (n=5), the anticommissural plication, Stanford modification, straight graft (SG), Uni-Graft, and Valsalva graft configurations were tested in series using an incomplete counterbalanced measures design, with the native root as a control, to mitigate ordering effects. Hemodynamic and videometric data were analyzed using linear models with conduit as the fixed effect of interest and valve as a fixed nuisance effect with post hoc pairwise testing using Tukey's correction.

RESULTS

Hemodynamics were clinically similar between grafts and control aortic roots. Regurgitant fraction varied between grafts, with SG and Uni-Graft groups having the lowest regurgitant fractions and anticommissural plication having the highest. Root distensibility was significantly lower in SG versus both control roots and all other grafts aside from the Stanford modification (P≤0.01 for each). All grafts except SG had significantly higher cusp opening velocities versus native roots (P<0.01 for each). Relative cusp opening forces were similar between SG, Uni-Graft, and control groups, whereas anticommissural plication, Stanford modification, and Valsalva grafts had significantly higher opening forces versus controls (P<0.01). Cusp closing velocities were similar between native roots and the SG group, and were significantly lower than observed in the other conduits (P≤0.01 for each). Only SG and Uni-Graft groups experienced relative cusp closing forces approaching that of the native root, whereas relative forces were >5-fold higher in the anticommissural plication, Stanford modification, and Valsalva graft groups.

CONCLUSIONS

In this ex vivo modeling system, clinically used valve-sparing aortic root replacement conduit configurations have comparable hemodynamics but differ in biomechanical performance, with the straight graft most closely recapitulating native aortic root biomechanics.

Для цитирования: Paulsen Michael J, Imbrie-Moore Annabel M, Baiocchi Michael, Wang Hanjay, Hironaka Camille E, Lucian Haley J, Farry Justin M, Thakore Akshara D, Zhu Yuanjia, Ma Michael, MacArthur John W, Woo Y Joseph. Comprehensive Ex Vivo Comparison of 5 Clinically Used Conduit Configurations for Valve-Sparing Aortic Root Replacement Using a 3-Dimensional-Printed Heart Simulator.. Circulation. 2020. DOI: 10.1161/circulationaha.120.046612.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.