DOI 10.1016/j.jtcvs.2020.06.028

Медицинская библиотека · The Journal of thoracic and cardiovascular surgery · 2020

Novel bicuspid aortic valve model with aortic regurgitation for hemodynamic status analysis using an ex vivo simulator.

Zhu Yuanjia, Imbrie-Moore Annabel M, Paulsen Michael J, Priromprintr Bryant, Wang Hanjay, Lucian Haley J, Farry Justin M, Woo Y Joseph

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАVideo-Audio MediaПеревод пока не подготовлен

Аннотация

OBJECTIVE

The objective was to design and evaluate a clinically relevant, novel ex vivo bicuspid aortic valve model that mimics the most common human phenotype with associated aortic regurgitation.

METHODS

Three bovine aortic valves were mounted asymmetrically in a previously validated 3-dimensional-printed left heart simulator. The non-right commissure and the non-left commissure were both shifted slightly toward the left-right commissure, and the left and right coronary cusps were sewn together. The left-right commissure was then detached and reimplanted 10 mm lower than its native height. Free margin shortening was used for valve repair. Hemodynamic status, high-speed videography, and echocardiography data were collected before and after the repair.

RESULTS

The bicuspid aortic valve model was successfully produced and repaired. High-speed videography confirmed prolapse of the fused cusp of the baseline bicuspid aortic valve models in diastole. Hemodynamic and pressure data confirmed accurate simulation of diseased conditions with aortic regurgitation and the subsequent repair. Regurgitant fraction postrepair was significantly reduced compared with that at baseline (14.5 ± 4.4% vs 28.6% ± 3.4%; P = .037). There was no change in peak velocity, peak gradient, or mean gradient across the valve pre- versus postrepair: 293.3 ± 18.3 cm/sec versus 325.3 ± 58.2 cm/sec (P = .29), 34.3 ± 4.2 mm Hg versus 43.3 ± 15.4 mm Hg (P = .30), and 11 ± 1 mm Hg versus 9.3 ± 2.5 mm Hg (P = .34), respectively.

CONCLUSIONS

An ex vivo bicuspid aortic valve model was designed that recapitulated the most common human phenotype with aortic regurgitation. These valves were successfully repaired, validating its potential for evaluating valve hemodynamics and optimizing surgical repair for bicuspid aortic valves.

Для цитирования: Zhu Yuanjia, Imbrie-Moore Annabel M, Paulsen Michael J, Priromprintr Bryant, Wang Hanjay, Lucian Haley J, Farry Justin M, Woo Y Joseph. Novel bicuspid aortic valve model with aortic regurgitation for hemodynamic status analysis using an ex vivo simulator.. The Journal of thoracic and cardiovascular surgery. 2020. DOI: 10.1016/j.jtcvs.2020.06.028.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.