DOI 10.1016/j.ijcard.2020.07.024

Медицинская библиотека · International journal of cardiology · 2020

Hemodynamic performance of the balloon-expandable SAPIEN 3 valve as assessed by cardiac magnetic resonance.

Nunes Ferreira-Neto Alfredo, Merten Constanze, Beurich Hans-Wilko, Zachow Dirk, Richardt Gert, Larose Eric, Guimaraes Leonardo, Pibarot Philippe, Pelletier-Beaumont Emilie, Rodés-Cabau Josep, Abdel-Wahab Mohamed

СтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Scarce data exist on transcatheter heart valve (THV) performance evaluated by cardiac magnetic resonance (CMR) in newer generation THV patients. Furthermore, it has been suggested that echocardiographic evaluation after TAVR may inaccurately assess residual AR in some patients. This study aimed to determine the incidence and severity of aortic regurgitation (AR) assessed by CMR in patients undergoing TAVR with the SAPIEN 3 valve, and evaluate the agreement between CMR and transthoracic echocardiography (TTE) on the assessment of AR severity in such patients.

METHODS

This multicentric observational study included 146 SAPIEN 3 patients with TTE and CMR within the month following their procedure. According to the CMR regurgitation fraction (RF), AR was considered mild and moderate-severe if the RF was 15-<30% and ≥ 30%, respectively. TTE exams followed VARC-2 recommendations.

RESULTS

By CMR, SAPIEN 3 recipients displayed a mean RF of 5.0 ± 6.1%, and mild and moderate-severe AR rates of 3.4% and 0.7%, respectively. The agreement between CMR-TTE was modest (weighted κ = 0.2640, p<0.001), due to an overestimation of AR severity by TTE. A historical cohort of 139 SAPIEN XT patients with a post-procedure CMR, displayed a mean RF of 9.6 ± 10.7% and mild and moderate-severe AR rates of 18.7% and 3.6%, respectively (p < .001 vs. SAPIEN 3 group).

CONCLUSIONS

SAPIEN 3 recipients exhibited very low rates of residual AR by CMR, suggesting a surgical-like performance regarding AR with this newer generation THV. TTE tended to overestimate the severity of AR, particularly among mild AR patients.

Для цитирования: Nunes Ferreira-Neto Alfredo, Merten Constanze, Beurich Hans-Wilko, Zachow Dirk, Richardt Gert, Larose Eric, Guimaraes Leonardo, Pibarot Philippe, Pelletier-Beaumont Emilie, Rodés-Cabau Josep, Abdel-Wahab Mohamed. Hemodynamic performance of the balloon-expandable SAPIEN 3 valve as assessed by cardiac magnetic resonance.. International journal of cardiology. 2020. DOI: 10.1016/j.ijcard.2020.07.024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.