[Multiple inappropriate defibrillator shocks due to insulation failure of a Biotronik Linox defibrillator lead with externalized conductor].
Клинический случайСтатьяПеревод пока не подготовлен
Аннотация
In this article the case of a patient who received a total of 35 inappropriate defibrillator shocks due to insulation failure with externalized conductor of a Biotronik Linox® lead is described. The implanted defibrillator was immediately inactivated and the failed lead was extracted using a laser sheath system.
Для цитирования: Elfarra Hamdi, Moosdorf Rainer, Rybinski Leszek, Grimm Wolfram. [Multiple inappropriate defibrillator shocks due to insulation failure of a Biotronik Linox defibrillator lead with externalized conductor].. Herzschrittmachertherapie & Elektrophysiologie. 2016. DOI: 10.1007/s00399-016-0418-6.