Twenty-four-Hour Intraocular Pressure-Related Patterns from Contact Lens Sensors in Normal-Tension Glaucoma and Healthy Eyes: The Exploring Nyctohemeral Intraocular pressure related pattern for Glaucoma Management (ENIGMA) Study.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Kim Yong Woo, Kim Jin-Soo, Lee Sang Yoon, Ha Ahnul, Lee Jinho, Park Young Joo, Kim Young Kook, Jeoung Jin Wook, Park Ki Ho
Аннотация
PURPOSE: To investigate 24-hour nyctohemeral intraocular pressure (IOP)-related patterns with contact lens sensors (CLSs) in eyes with primary open-angle glaucoma (POAG) with normal baseline IOP (i.e., normal-tension glaucoma [NTG]) and healthy controls.
DESIGN: Prospective, case-control study.
PARTICIPANTS: Thirty eyes of 30 patients with NTG, who had had a wash-out period for their IOP-lowering treatment, and 20 eyes of 20 healthy volunteer subjects.
METHODS: Patients and subjects were hospitalized for the purposes of 24-hour CLS (SENSIMED Triggerfish; Sensimed AG, Lausanne, Switzerland) measurement. The IOP-related patterns during wake and sleep times over the course of the 24 hours were compared between the 2 groups. The 24-hour ambulatory blood pressure and posture were monitored simultaneously. A generalized linear model was used to find the factors associated with NTG.
MAIN OUTCOME MEASURES: The IOP-related patterns, including mean and standard deviation (SD) of measurements, amplitude of cosine-fit curve, acrophase (signal peak), and bathyphase (signal trough) values (millivolt equivalents [mVEq]).
RESULTS: The SDs of the 24-hour CLS measurements were significantly greater in NTG eyes than in healthy controls (112.51±26.90 vs. 85.18±29.61 mVEq, P = 0.002). The amplitudes of cosine-fit curve (141.88±39.96 vs. 106.08±41.49 mVEq, P = 0.004) and acrophase values (277.74±129.80 vs. 190.58±127.88 mVEq, P = 0.024), mostly measured during nocturnal period, were significantly greater in NTG eyes than in healthy controls. The NTG subjects slept longer in the lateral decubitus posture than the healthy controls (199.1±137.8 vs. 113.2±86.2 minutes, P = 0.009). In the multivariable generalized linear model, the greater amplitude of cosine-fit curve (β = 0.218, P = 0.012) and greater time of decubitus posture during sleep (β = 0.180, P = 0.004) were found to be significantly associated with NTG.
CONCLUSIONS: Continuous monitoring of 24-hour IOP-related values with CLS can be useful for assessment of glaucoma risk, especially for patients with NTG whose IOP appears to be in the normal range. Fluctuation of 24-hour IOP-related values and posture during sleep time might be associated with NTG.
DESIGN: Prospective, case-control study.
PARTICIPANTS: Thirty eyes of 30 patients with NTG, who had had a wash-out period for their IOP-lowering treatment, and 20 eyes of 20 healthy volunteer subjects.
METHODS: Patients and subjects were hospitalized for the purposes of 24-hour CLS (SENSIMED Triggerfish; Sensimed AG, Lausanne, Switzerland) measurement. The IOP-related patterns during wake and sleep times over the course of the 24 hours were compared between the 2 groups. The 24-hour ambulatory blood pressure and posture were monitored simultaneously. A generalized linear model was used to find the factors associated with NTG.
MAIN OUTCOME MEASURES: The IOP-related patterns, including mean and standard deviation (SD) of measurements, amplitude of cosine-fit curve, acrophase (signal peak), and bathyphase (signal trough) values (millivolt equivalents [mVEq]).
RESULTS: The SDs of the 24-hour CLS measurements were significantly greater in NTG eyes than in healthy controls (112.51±26.90 vs. 85.18±29.61 mVEq, P = 0.002). The amplitudes of cosine-fit curve (141.88±39.96 vs. 106.08±41.49 mVEq, P = 0.004) and acrophase values (277.74±129.80 vs. 190.58±127.88 mVEq, P = 0.024), mostly measured during nocturnal period, were significantly greater in NTG eyes than in healthy controls. The NTG subjects slept longer in the lateral decubitus posture than the healthy controls (199.1±137.8 vs. 113.2±86.2 minutes, P = 0.009). In the multivariable generalized linear model, the greater amplitude of cosine-fit curve (β = 0.218, P = 0.012) and greater time of decubitus posture during sleep (β = 0.180, P = 0.004) were found to be significantly associated with NTG.
CONCLUSIONS: Continuous monitoring of 24-hour IOP-related values with CLS can be useful for assessment of glaucoma risk, especially for patients with NTG whose IOP appears to be in the normal range. Fluctuation of 24-hour IOP-related values and posture during sleep time might be associated with NTG.