Eckel Clemens, Kim Won-Keun, Sötemann Dagmar, Grothusen Christina, Tiyerili Vedat, Dohmen Guido, Renker Matthias, Charitos Efstratios I, Hamm Christian W, Choi Yeong-Hoon, Elsässer Albrecht, Möllmann Helge, Blumenstein Johannes
Аннотация
BACKGROUND: Comparative data on transcatheter self-expanding ACURATE neo2 (NEO2) and balloon-expandable SAPIEN 3 Ultra prostheses in technically challenging anatomy of severe aortic valve calcified aortic annuli are scarce.
METHODS: A total of 1987 patients with severe native aortic stenosis treated with the self-expanding NEO2 (n=1457) or balloon-expandable SAPIEN 3 Ultra (n=530) from January 2017 to April 2023 were evaluated. The primary end point was procedural outcome according to the Valve Academic Research Consortium 3 definitions. Propensity matching defined 219 pairs with severe calcification (calcium density cutoff, 758 AU/cm2) of the native aortic valve.
RESULTS: Technical success (90.4% versus 91.8%; risk difference, 1.4% [95% CI, -4.4 to -7.2]; P=0.737) and device success at 30 days (80.8% versus 75.8%; risk difference, -5.0% [95% CI, -13.2 to 3.1]; P=0.246) were comparable between NEO2 and SAPIEN 3 Ultra. The rate of severe prosthesis-patient mismatch (1.1% versus 10.1%; risk difference, 10.0% [95% CI, 4.0-13.9]; P<0.001) and mean transvalvular gradient ≥20 mm Hg (2.8% versus 14.3%; risk difference, 11.5% [95% CI, 5.8-17.1]; P<0.001) was lower with NEO2. The rate of more-than-mild paravalvular leakage or valve-in-valve due to paravalvular leakage was significantly higher (6.2% versus 0.0%; risk difference, 6.2% [95% CI, -10.1 to -2.7]; P=0.002), and there was a tendency for a higher rate of device embolization or migration (1.8% versus 0.0%; risk difference, -1.8% [95% CI, -4.1 to 0.4]; P=0.123) with NEO2. Multivarate regression revealed no independent impact of transcatheter heart valve selection on device success (odds ratio, 0.93 [95% CI, 0.48-1.77]; P=0.817).
CONCLUSIONS: In patients with severely calcified annuli, supraannular implantation of NEO2 showed hemodynamic advantages. Nevertheless, NEO2 was associated with a higher incidence of relevant paravalvular leakage and a numerically higher rate of device embolization than SAPIEN 3 Ultra in this particular patient group.
METHODS: A total of 1987 patients with severe native aortic stenosis treated with the self-expanding NEO2 (n=1457) or balloon-expandable SAPIEN 3 Ultra (n=530) from January 2017 to April 2023 were evaluated. The primary end point was procedural outcome according to the Valve Academic Research Consortium 3 definitions. Propensity matching defined 219 pairs with severe calcification (calcium density cutoff, 758 AU/cm2) of the native aortic valve.
RESULTS: Technical success (90.4% versus 91.8%; risk difference, 1.4% [95% CI, -4.4 to -7.2]; P=0.737) and device success at 30 days (80.8% versus 75.8%; risk difference, -5.0% [95% CI, -13.2 to 3.1]; P=0.246) were comparable between NEO2 and SAPIEN 3 Ultra. The rate of severe prosthesis-patient mismatch (1.1% versus 10.1%; risk difference, 10.0% [95% CI, 4.0-13.9]; P<0.001) and mean transvalvular gradient ≥20 mm Hg (2.8% versus 14.3%; risk difference, 11.5% [95% CI, 5.8-17.1]; P<0.001) was lower with NEO2. The rate of more-than-mild paravalvular leakage or valve-in-valve due to paravalvular leakage was significantly higher (6.2% versus 0.0%; risk difference, 6.2% [95% CI, -10.1 to -2.7]; P=0.002), and there was a tendency for a higher rate of device embolization or migration (1.8% versus 0.0%; risk difference, -1.8% [95% CI, -4.1 to 0.4]; P=0.123) with NEO2. Multivarate regression revealed no independent impact of transcatheter heart valve selection on device success (odds ratio, 0.93 [95% CI, 0.48-1.77]; P=0.817).
CONCLUSIONS: In patients with severely calcified annuli, supraannular implantation of NEO2 showed hemodynamic advantages. Nevertheless, NEO2 was associated with a higher incidence of relevant paravalvular leakage and a numerically higher rate of device embolization than SAPIEN 3 Ultra in this particular patient group.
Предметные рубрики MeSH
HumansAortic Valve StenosisMaleHeart Valve ProsthesisFemaleAortic ValveProsthesis DesignTranscatheter Aortic Valve ReplacementAgedAged, 80 and overSeverity of Illness IndexCalcinosisTreatment OutcomeRisk FactorsTime FactorsRisk AssessmentRetrospective StudiesPropensity ScoreRecovery of FunctionBalloon ValvuloplastyHemodynamics