DOI 10.1177/11206721211020633

Медицинская библиотека · European journal of ophthalmology · 2021

Comparison of total keratometry with corneal power measured by optical low-coherence reflectometry and placido-dual Scheimpflug system.

Kommineni Usha Bhanu, Mohamed Ashik, Vaddavalli Pravin K, Reddy Jagadesh C

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the total keratometry (TK) and astigmatism measurements in eyes with cataract using automated keratometry of swept-source optical coherence tomography (ss-OCT), optical low-coherence reflectometry (OLCR), simulated keratometry (SimK), and total corneal power (TCP) of combined placido-dual Scheimpflug imaging system.

SETTING

The study was conducted at LV Prasad Eye Institute, Hyderabad, India.

DESIGN

Retrospective evaluation of electronic medical records of patients who were evaluated for cataract surgery.

METHODS

Twenty-eight eyes of 28 patients were included in the study. All patients evaluated for cataract surgery underwent corneal power measurements using three devices: ssOCT, OLCR, and combined placido-dual Scheimpflug imaging were included in the study. Vector analysis was performed to evaluate corneal astigmatism and Bland-Altman analysis was conducted to evaluate the limits of agreement of similar parameters among devices.

RESULTS

The mean TK was statistically significantly different from the keratometry obtained from optical biometers and values measured by the Scheimpflug imaging system. The magnitude of mean difference was greater between TK and TCP (0.75 ± 0.25) compared to other variables. The mean difference in astigmatism between TK, ss-OCT-K (0.09 ± 0.12, p = 0.48), OCLR-K (0.10 ± 0.48, p = 0.91), and TCP (0.09 ± 0.47, p = 0.31) was not statistically significant but was statistically significant between TK and SimK values (0.23D ± 0.49D). The axis of orientation (<20°) of astigmatism was comparable (100%, 28 eyes) between two keratometry variables measured by ss-OCT.

CONCLUSION

There appears to be a greater correlation of automated keratometry, and TK values obtained from ss-OCT compared to other variables studied. The measurements from TK, Simk, and TCP cannot be used interchangeably.

Для цитирования: Kommineni Usha Bhanu, Mohamed Ashik, Vaddavalli Pravin K, Reddy Jagadesh C. Comparison of total keratometry with corneal power measured by optical low-coherence reflectometry and placido-dual Scheimpflug system.. European journal of ophthalmology. 2021. DOI: 10.1177/11206721211020633.