MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.juro.2012.08.244

Медицинская библиотека · The Journal of urology · 2012

Masked pneumothorax: risk of valveless trocar systems.

Hillelsohn Joel H, Friedlander Justin I, Bagadiya Neeti, Okhunov Zhamshid, Kashan Mahyar, Schwartzur Mikhail, Kavoussi Louis

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Unlike traditional valved trocars, the valveless trocar maintains pneumoperitoneum during laparoscopy by forming a CO(2) curtain at the proximal end of the trocar. This gas barrier instantaneously maintains exact intraperitoneal pressure that yields to the transient physiological changes seen with breathing. Due to this different mechanism of action, pneumothorax development may be masked by the valveless trocar system.

MATERIALS AND

METHODS

We retrospectively reviewed 850 transperitoneal laparoscopic kidney and adrenal surgeries in which a valveless trocar system was used to determine any record of pneumothorax detected intraoperatively or postoperatively. A patient with pneumothorax was considered a case and anesthetic parameters were reviewed. A matched control group was generated from patients treated with transperitoneal laparoscopic kidney and adrenal surgery using the valveless trocar with no complications.

RESULTS

Pneumothorax was diagnosed in 10 patients (1.2%). Two cases were the result of intentional excision of the diaphragm, which were repaired intraoperatively, while 8 were not recognized until the postoperative period. Five of the patients (63%) with unintentional pneumothorax required chest tube placement for a mean of 2.4 days. The remaining 3 patients (37%) were treated conservatively and followed with serial chest x-rays. The only anesthetic variable that was significantly different between the groups was Δ end tidal CO(2) with greater fluctuations in end tidal CO(2) in the pneumothorax group than in controls (p = 0.03).

CONCLUSIONS

Pneumothorax is a rare complication of laparoscopic urological surgery that is usually recognized intraoperatively through physiological changes. Valveless trocar systems mask these findings and can delay identification until the postoperative period.

Для цитирования: Hillelsohn Joel H, Friedlander Justin I, Bagadiya Neeti, Okhunov Zhamshid, Kashan Mahyar, Schwartzur Mikhail, Kavoussi Louis. Masked pneumothorax: risk of valveless trocar systems.. The Journal of urology. 2012. DOI: 10.1016/j.juro.2012.08.244.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление