Electrogram morphology discriminators in implantable cardioverter defibrillators: A comparative evaluation.
Сравнительное исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Frontera Antonio, Strik Marc, Eschalier Romain, Biffi Mauro, Pereira Bruno, Welte Nicolas, Chauvel Remi, Mondoly Pierre, Laborderie Julien, Bernis Jean-Paul, Clementy Nicolas, Reuter Sylvain, Garrigue Stéphane, Deplagne Antoine, Vernooy Kevin, Pillois Xavier, Haïssaguerre Michel, Dubois Remi, Ritter Philippe, Bordachar Pierre, Ploux Sylvain
Аннотация
BACKGROUND: Morphology algorithms are currently recommended as a standalone discriminator in single-chamber implantable cardioverter defibrillators (ICDs). However, these proprietary algorithms differ in both design and nominal programming.
OBJECTIVE: To compare three different algorithms with nominal versus advanced programming in their ability to discriminate between ventricular (VT) and supraventricular tachycardia (SVT).
METHODS: In nine European centers, VT and SVTs were collected from Abbott, Boston Scientific, and Medtronic dual- and triple-chamber ICDs via their respective remote monitoring portals. Percentage morphology matches were recorded for selected episodes which were classified as VT or SVT by means of atrioventricular comparison. The sensitivity and related specificity of each manufacturer discriminator was determined at various values of template match percentage from receiving operating characteristics (ROC) curve analysis.
RESULTS: A total of 534 episodes were retained for the analysis. In ROC analyses, Abbott Far Field MD (area under the curve [AUC]: 0.91; P < .001) and Boston Scientific RhythmID (AUC: 0.95; P < .001) show higher AUC than Medtronic Wavelet (AUC: 0.81; P < .001) when tested for their ability to discriminate VT from SVT. At nominal % match threshold all devices provided high sensitivity in VT identification, (91%, 100%, and 90%, respectively, for Abbott, Boston Scientific, and Medtronic) but contrasted specificities in SVT discrimination (85%, 41%, and 62%, respectively). Abbott and Medtronic's nominal thresholds were similar to the optimal thresholds. Optimization of the % match threshold improved the Boston Scientific specificity to 79% without compromising the sensitivity.
CONCLUSION: Proprietary morphology discriminators show important differences in their ability to discriminate SVT. How much this impact the overall discrimination process remains to be investigated.
OBJECTIVE: To compare three different algorithms with nominal versus advanced programming in their ability to discriminate between ventricular (VT) and supraventricular tachycardia (SVT).
METHODS: In nine European centers, VT and SVTs were collected from Abbott, Boston Scientific, and Medtronic dual- and triple-chamber ICDs via their respective remote monitoring portals. Percentage morphology matches were recorded for selected episodes which were classified as VT or SVT by means of atrioventricular comparison. The sensitivity and related specificity of each manufacturer discriminator was determined at various values of template match percentage from receiving operating characteristics (ROC) curve analysis.
RESULTS: A total of 534 episodes were retained for the analysis. In ROC analyses, Abbott Far Field MD (area under the curve [AUC]: 0.91; P < .001) and Boston Scientific RhythmID (AUC: 0.95; P < .001) show higher AUC than Medtronic Wavelet (AUC: 0.81; P < .001) when tested for their ability to discriminate VT from SVT. At nominal % match threshold all devices provided high sensitivity in VT identification, (91%, 100%, and 90%, respectively, for Abbott, Boston Scientific, and Medtronic) but contrasted specificities in SVT discrimination (85%, 41%, and 62%, respectively). Abbott and Medtronic's nominal thresholds were similar to the optimal thresholds. Optimization of the % match threshold improved the Boston Scientific specificity to 79% without compromising the sensitivity.
CONCLUSION: Proprietary morphology discriminators show important differences in their ability to discriminate SVT. How much this impact the overall discrimination process remains to be investigated.
Предметные рубрики MeSH
Action PotentialsAlgorithmsDefibrillators, ImplantableDiagnosis, DifferentialElectric CountershockElectrophysiologic Techniques, CardiacEquipment DesignEuropeHeart RateHumansPredictive Value of TestsReproducibility of ResultsRetrospective StudiesSignal Processing, Computer-AssistedTachycardia, SupraventricularTachycardia, VentricularTelemetry