Unique technique to relieve left ventricular assist device electromagnetic interference with an implantable cardioverter defibrillator.

Jin Chengyue, Hsu Joshua, Frenkel Daniel, Jacobson Jason T, Iwai Sei, Ferrick Aileen

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

We introduced a simple technique to eliminate electromagnetic interference between a left ventricular assist device (LVAD) and an implantable cardioverter defibrillator (ICD). A 43-year-old male with heart failure and a reduced ejection fraction who had an ICD presented with decompensated heart failure and received an LVAD as a bridge to transplant. Remote monitoring showed persistent atrial fibrillation causing an inappropriate ICD shock leading to a decision to disable shock therapies. However, an in-office interrogation was unsuccessful due to electromagnetic interference. Patient was instructed to extend his arm above his head on the ipsilateral side of the ICD, thus increasing the distance between LVAD and ICD, eliminating the interaction to allow reprogramming of the device.

Для цитирования: Jin Chengyue, Hsu Joshua, Frenkel Daniel, Jacobson Jason T, Iwai Sei, Ferrick Aileen. Unique technique to relieve left ventricular assist device electromagnetic interference with an implantable cardioverter defibrillator.. Journal of cardiovascular electrophysiology. 2021. DOI: 10.1111/jce.14840.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.