Медицинская библиотека · Journal of cataract and refractive surgery · 2011

Randomized comparison of postoperative use of hydrogel ocular bandage and collagen corneal shield for wound protection and patient tolerability after cataract surgery.

Dell Steven J, Hovanesian John A, Raizman Michael B, Crandall Alan S, Doane John, Snyder Michael, Masket Samuel, Lane Stephen, Fram Nicole, Ocular Bandage Study Group

Сравнительное исследованиеСтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the safety and efficacy of a hydrogel bandage and a collagen corneal shield in providing wound protection and relief of pain/discomfort in the acute period after uneventful unilateral clear corneal phacoemulsification cataract surgery with foldable intraocular lens (IOL) implantation.

SETTING

Seventeen investigational sites in the United States.

DESIGN

Prospective randomized single-masked parallel study.

METHODS

The study comprised patients scheduled to have unilateral clear corneal cataract surgery with posterior chamber intraocular lens implantation. The patients were examined preoperatively and frequently for 30 days postoperatively. The design was a noninferiority study of the 2 primary endpoints, device performance and maximum reported postoperative pain.

RESULTS

The device performance success was 78.6% (228/290) for the hydrogel bandage and 26.5% (26/98) for the corneal shield (P<.0001 for noninferiority). Analyses indicated that the hydrogel bandage was superior to the corneal shield in device performance (P<.001; difference = 52.1%; 95% confidence interval, 41.6%-61.4%). The maximum postoperative pain/discomfort score of the hydrogel bandage (mean 1.3 ± 1.8 [SD]; scale 0 to 10) was noninferior to that of the corneal shield (1.1 ± 1.6) in the first 4 hours after surgery (P<.001). Adverse events in the cataract surgeries were reported in 22.2% (70/316) and 36.5% (38/104) of hydrogel bandage patients and corneal shield patients, respectively (P = .0045).

CONCLUSION

The hydrogel bandage was safe and effective for ocular surface protection and relief of pain/discomfort when applied topically to clear corneal incisions used in cataract or IOL implantation surgery.

FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned. Additional disclosures are found in the footnotes.

Для цитирования: Dell Steven J, Hovanesian John A, Raizman Michael B, Crandall Alan S, Doane John, Snyder Michael, Masket Samuel, Lane Stephen, Fram Nicole, Ocular Bandage Study Group. Randomized comparison of postoperative use of hydrogel ocular bandage and collagen corneal shield for wound protection and patient tolerability after cataract surgery.. Journal of cataract and refractive surgery. 2011. DOI: 10.1016/j.jcrs.2010.11.001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.