Evaluation of a Novel Field Suction Unit (SÜPEVAC): A Mannequin Crossover Trial.
СтатьяРандомизированное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Baxter Teneal, Grimmett Wallace, Reardon Damon
Аннотация
OBJECTIVE: The purpose of this study was to obtain a quantitative and qualitative assessment of the SÜPEVAC unit (German Malaysian Medical Industries, Selangor, Malaysia), a novel field suction unit powered by a positive pressure gas source and capable of hospital-grade suction levels and rates.
METHODS: This was a randomized crossover trial in a simulated clinical setting to compare the SÜPEVAC unit with standard hospital wall suction. Twenty-two fellows of the Australasian College for Emergency Medicine, the Australian and New Zealand College of Anaesthetists, or the College of Intensive Care Medicine were recruited. Outcomes were assessed via a structured survey and measuring the time to view the glottis.
RESULTS: This study found there was no significant difference between the SÜPEVAC unit and wall suction with regard to suction time (P = .762; 95% confidence interval, -0.683 to 0.338) or qualitative assessment via the survey.
CONCLUSIONS: The SÜPEVAC unit is comparable with wall suction in a clinical setting. Further research is warranted.
METHODS: This was a randomized crossover trial in a simulated clinical setting to compare the SÜPEVAC unit with standard hospital wall suction. Twenty-two fellows of the Australasian College for Emergency Medicine, the Australian and New Zealand College of Anaesthetists, or the College of Intensive Care Medicine were recruited. Outcomes were assessed via a structured survey and measuring the time to view the glottis.
RESULTS: This study found there was no significant difference between the SÜPEVAC unit and wall suction with regard to suction time (P = .762; 95% confidence interval, -0.683 to 0.338) or qualitative assessment via the survey.
CONCLUSIONS: The SÜPEVAC unit is comparable with wall suction in a clinical setting. Further research is warranted.