Медицинская библиотека · Anesthesia, essays and researches · 2021

A Comparative Study between Truview PCD Video Laryngoscope and Macintosh Laryngoscope with Respect to Intubation Quality and Hemodynamic Changes.

Samal Rajiba Lochan, Swain Sumita, Samal Soumya

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND AIMS: Video laryngoscopes resemble traditional laryngoscopes, but they have a video chip embedded in the tip of laryngoscope blade. This enables the operator to "look around the corners" which is not possible with conventional direct laryngoscopes. The present study was undertaken to compare Truview video laryngoscope and Macintosh laryngoscope for glottis visualization, ease of tracheal intubation, and associated hemodynamic response.

SETTING

The study was conducted in operation theater in a medical college.

STUDY

DESIGN

It was a randomized prospective observational study.

MATERIALS AND

METHODS

Sixty patients of American Society of Anesthesiologists Grade 1 and 2 of either sex aged 18-60 years who were scheduled to undergo elective surgery requiring general anesthesia with orotracheal intubation were selected. In patients of Group T (n = 30), intubation was done using Truview video laryngoscope, while in Group M (n = 30), intubation was done using Macintosh laryngoscope. Various airway and hemodynamic parameters were assessed and compared.

STATISTICAL ANALYSIS: Statistical analysis was done using Chi-square test, paired and unpaired Student's t-test, and ANOVA test. P < 0.05 is considered statistically significant.

RESULTS

Distribution of modified Mallampati Class (MMPC), ease of laryngoscopic blade insertion, and size of cuffed endotracheal tube used were statistically comparable in both the groups. The time to intubation was more in Group T (37.16 ± 8.23 s) as compared to Group M (29.80 ± 6.75 s). There was a statistically significant better modified Cormack and Lehane (CL) grading view obtained in Group T as compared to Group M (P = 0.025). CL Grades 2 and 3 were not seen in any of either of the group. The mean intubation difficulty score (IDS) was significantly lower in Group T (0.3 ± 0.60) as compared to Group M (0.73 ± 0.86). In both the Groups T and M, the mean heart rate, systolic blood pressure (BP), and diastolic BP were significantly increased from baseline for up to 3 min after laryngoscopy, but they were comparable between the two groups all the time.

CONCLUSION

Truview propaganda cum distribution laryngoscope provides a better glottis view than the Macintosh laryngoscope. Although it requires a longer time to intubate using Truview, the overall IDS score was lower as compared to Macintosh laryngoscope. Hemodynamic changes remained similar in both the groups.

Для цитирования: Samal Rajiba Lochan, Swain Sumita, Samal Soumya. A Comparative Study between Truview PCD Video Laryngoscope and Macintosh Laryngoscope with Respect to Intubation Quality and Hemodynamic Changes.. Anesthesia, essays and researches. 2021. DOI: 10.4103/aer.aer_62_21.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.