Diagnosing complex regional pain syndrome using an electronic nose, a pilot study.

Bijl E J, Groeneweg J G, Wesselius D W, Stronks D L, Huygen F J P M

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Complex regional pain syndrome (CRPS) is a complication after surgery or trauma and is characterized by a continuing regional pain in a distal extremity. The pain is disproportionate in severity and duration in relation to the preceding trauma. Currently, the diagnosis is based on the patients' signs and symptoms. There is no objective clinically applicable test available to confirm the diagnosis of CRPS, however this could contribute to a more reliable and valid diagnosis. Since the treatment of CRPS differs from that of other types of pain this could thereby lead to earlier and (more) appropriate treatment and possibly to lower medical costs. The Aeonose™ is a diagnostic test device which detects volatile organic profiles in exhaled air. Exhaled breath analysis using an electronic nose has been successfully applied to differentiate between sick and healthy persons for various indications. This study was a feasibility study in which we investigated whether the Aeonose™ is able to measure a difference in the volatome of CRPS patients compared to the volatome of healthy controls.

DESIGN

Prospective observational study.

SETTING

University Center for Pain Medicine.

SUBJECTS: Adult patients diagnosed with CRPS according to the latest IASP criteria (n = 36) and matched healthy controls (n = 36).

METHODS

Breath profiles were sampled by breathing in and out through the Aeonose™. Data were compressed using a Tucker3-like solution and subsequently used for training an artificial neural network together with the classification 'CRPS: Yes' or 'CRPS: No'. Cross-validation was applied using the leave-10%-out method.

RESULTS

Data of the 72 participants were analyzed, resulting in a sensitivity of 83% (95% CI 67%-93%), specificity of 78% (95% CI 60%-89%), and an overall accuracy of 81%.

CONCLUSIONS

This study suggests that the Aeonose™ can possibly distinguish patients with CRPS from healthy controls based on analysis of their volatome (MEC-2014-149).

Для цитирования: Bijl E J, Groeneweg J G, Wesselius D W, Stronks D L, Huygen F J P M. Diagnosing complex regional pain syndrome using an electronic nose, a pilot study.. Journal of breath research. 2019. DOI: 10.1088/1752-7163/aaf9c1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.