MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1089/pho.2015.3993

Медицинская библиотека · Photomedicine and laser surgery · 2016

Descriptive Analysis of In Vitro Cutting of Swine Mitral Cusps: Comparison of High-Power Laser and Scalpel Blade Cutting Techniques.

Pinto Nathali Cordeiro, Pomerantzeff Pablo Maria Alberto, Deana Alessandro, Zezell Denise, Benetti Carolina, Aiello Vera Demarchi, Lopes Luciana Almeida, Jatene Fabio Biscegli, Chavantes M Cristina

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND

OBJECTIVES

The most common injury to the heart valve with rheumatic involvement is mitral stenosis, which is the reason for a big number of cardiac operations in Brazil. Commissurotomy is the traditional technique that is still widely used for this condition, although late postoperative restenosis is concerning. This study's purpose was to compare the histological findings of porcine cusp mitral valves treated in vitro with commissurotomy with a scalpel blade to those treated with high-power laser (HPL) cutting, using appropriate staining techniques.

MATERIALS AND

METHODS

Five mitral valves from healthy swine were randomly divided into two groups: Cusp group (G1), cut with a scalpel blade (n = 5), and Cusp group (G2), cut with a laser (n = 5). G2 cusps were treated using a diode laser (λ = 980 nm, power = 9.0 W, time = 12 sec, irradiance = 5625 W/cm2, and energy = 108 J).

RESULTS

In G1, no histological change was observed in tissue. A hyaline basophilic aspect was focally observed in G2, along with a dark red color on the edges and areas of lower birefringence, when stained with hematoxylin-eosin, Masson's trichrome, and Sirius red. Further, the mean distances from the cutting edge in cusps submitted to laser application and stained with Masson's trichrome and Sirius red were 416.7 and 778.6 μm, respectively, never overcoming 1 mm in length.

CONCLUSIONS

Thermal changes were unique in the group submitted to HPL and not observed in the cusp group cut with a scalpel blade. The mean distance of the cusps' collagen injury from the cutting edge was less than 1 mm with laser treatment. Additional studies are needed to establish the histological evolution of the laser cutting and to answer whether laser cutting may avoid valvular restenosis better than blade cutting.

Для цитирования: Pinto Nathali Cordeiro, Pomerantzeff Pablo Maria Alberto, Deana Alessandro, Zezell Denise, Benetti Carolina, Aiello Vera Demarchi, Lopes Luciana Almeida, Jatene Fabio Biscegli, Chavantes M Cristina. Descriptive Analysis of In Vitro Cutting of Swine Mitral Cusps: Comparison of High-Power Laser and Scalpel Blade Cutting Techniques.. Photomedicine and laser surgery. 2016. DOI: 10.1089/pho.2015.3993.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.