DOI 10.7754/clin.lab.2021.210316

Медицинская библиотека · Clinical laboratory · 2021

Rapid Cobas Liat SARS-CoV-2 Assay in Comparison with the Laboratory-Developed Real-Time RT-PCR Test.

Er Tze-Kiong, Chou Yung-Che, Chen Si-Yu, Huang Jian-Wei

Письмо в редакциюПеревод пока не подготовлен

Аннотация

BACKGROUND

The objective of this study was to compare the validity of two different assays for the detection of SARS-CoV-2.

METHODS

We collected 50 nasopharyngeal swabs in universal transport medium from the emergency department of Asia University Hospital for the detection of SARS-CoV-2 using reverse transcription-polymerase chain reaction (RT-PCR). The samples for the Liat SARS-CoV-2 influenza A/B test were stored at -70℃ after SARS-CoV-2 testing using the RT-PCR in order to assess method comparison.

RESULTS

In this study, the Limit of detection (LOD) of the cobas Liat SARS-CoV-2 and influenza A/B nucleic acid test is 12 copies/µL and the assay obtained 100% positive agreement and negative percent agreement with RT-PCR.

CONCLUSIONS

In summary, a prefect agreement exists between the detection of SARS-CoV-2 conducted with the cobas Liat SARS-CoV-2 and influenza A/B nucleic acid test and the RT-PCR. The cobas Liat SARS-CoV-2 and influenza A/B nucleic acid test is a reliable method for the detection of SARS-CoV-2, and it only requires 20 minutes to obtain the results. On the other hand, the cobas Liat SARS-CoV-2 and influenza A/B nucleic acid test is accurate, easy to use, and provides a faster turnaround time than testing performed in the high-throughput platform.

Для цитирования: Er Tze-Kiong, Chou Yung-Che, Chen Si-Yu, Huang Jian-Wei. Rapid Cobas Liat SARS-CoV-2 Assay in Comparison with the Laboratory-Developed Real-Time RT-PCR Test.. Clinical laboratory. 2021. DOI: 10.7754/clin.lab.2021.210316.