Медицинская библиотека · Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology · 2015

High performances of (18)F-fluorodeoxyglucose PET-CT in cardiac implantable device infections: A study of 40 patients.

Tlili Ghoufrane, Amraoui Sana, Mesguich Charles, Rivière Annalisa, Bordachar Pierre, Hindié Elif, Bordenave Laurence

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Cardiovascular implantable electronic devices (CIED) infections are life-threatening complications. The diagnosis can be difficult to establish. Our purpose is to evaluate the diagnostic value of (18)F-FDG PET.

METHODS

Forty patients who received work-up for suspected CIED infection were retrospectively included (group 1) and compared with 40 controls (group 2); CIED patients were referred for oncologic PET. PET-CT data were blindly assessed. Interpretation was based on visual analysis of both attenuation-corrected and non-corrected images and a semi-quantitative analysis was performed. The gold standard was bacteriological data of explanted devices or clinical follow-up for at least 1 year.

RESULTS

Infection was present in 18 out of 40 patients of group 1. Sensitivity, specificity, positive predictive value, and negative predictive value of PET-CT were 83%, 95%, 94%, and 88%, respectively. Accuracy was 90%. PET-CT revealed the presence of additional pathological hypermetabolic foci in 28% of cases. PET-CT was negative at implanted devices in all patients of group 2.

CONCLUSION

(18)F-FDG PET-CT is helpful in the work-up of suspected CIED infections. It is a potential tool to make the accurate diagnosis of CIED infection and to assess the extent of infection. The promising results in this indication need to be validated in a prospective multicenter study.

Для цитирования: Tlili Ghoufrane, Amraoui Sana, Mesguich Charles, Rivière Annalisa, Bordachar Pierre, Hindié Elif, Bordenave Laurence. High performances of (18)F-fluorodeoxyglucose PET-CT in cardiac implantable device infections: A study of 40 patients.. Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology. 2015. DOI: 10.1007/s12350-015-0067-x.