Radiofrequency cardiac ablation with catheters placed on opposing sides of the ventricular wall: computer modelling comparing bipolar and unipolar modes.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
González-Suárez Ana, Trujillo Macarena, Koruth Jacob, d'Avila Andre, Berjano Enrique
Аннотация
PURPOSE: The aim of this study was to compare the efficacy of bipolar (BM) vs. unipolar (UM) mode of radiofrequency ablation (RFA) in terms of creating transmural lesions across the interventricular septum (IVS) and ventricular free wall (VFW).
MATERIALS AND METHODS: We built computational models to study the temperature distributions and lesion dimensions created by BM and UM on IVS and VFW during RFA. Two different UM types were considered: sequential (SeUM) and simultaneous (SiUM). The effect of ventricular wall thickness, catheter misalignment, epicardial fat, and presence of air in the epicardial space were also studied.
RESULTS: Regarding IVS ablation, BM created transmural and symmetrical lesions for wall thicknesses up to 15 mm. SeUM and SiUM were not able to create transmural lesions with IVS thicknesses ≥12.5 and 15 mm, respectively. Lesions were asymmetrical only with SeUM. For VFW ablation, BM also created transmural lesions for wall thicknesses up to 15 mm. However, with SeUM and SiUM transmurality was obtained for VFW thicknesses ≤7.5 and 12.5 mm, respectively. With the three modes, VFW lesions were always asymmetrical. In the scenario with air or a fat tissue layer on the epicardial side, only SiUM was capable of creating transmural lesions. Overall, BM was superior to UM in IVS and VFW ablation when the catheters were not aligned.
CONCLUSIONS: Our findings suggest that BM is more effective than UM in achieving transmurality across both ventricular sites, except in the situation of the epicardial catheter tip surrounded by air or placed over a fat tissue layer.
MATERIALS AND METHODS: We built computational models to study the temperature distributions and lesion dimensions created by BM and UM on IVS and VFW during RFA. Two different UM types were considered: sequential (SeUM) and simultaneous (SiUM). The effect of ventricular wall thickness, catheter misalignment, epicardial fat, and presence of air in the epicardial space were also studied.
RESULTS: Regarding IVS ablation, BM created transmural and symmetrical lesions for wall thicknesses up to 15 mm. SeUM and SiUM were not able to create transmural lesions with IVS thicknesses ≥12.5 and 15 mm, respectively. Lesions were asymmetrical only with SeUM. For VFW ablation, BM also created transmural lesions for wall thicknesses up to 15 mm. However, with SeUM and SiUM transmurality was obtained for VFW thicknesses ≤7.5 and 12.5 mm, respectively. With the three modes, VFW lesions were always asymmetrical. In the scenario with air or a fat tissue layer on the epicardial side, only SiUM was capable of creating transmural lesions. Overall, BM was superior to UM in IVS and VFW ablation when the catheters were not aligned.
CONCLUSIONS: Our findings suggest that BM is more effective than UM in achieving transmurality across both ventricular sites, except in the situation of the epicardial catheter tip surrounded by air or placed over a fat tissue layer.