Electromagnetic interference of magnetic field based auto identification technologies in healthcare settings.

Kapa Suraj, Pierce Timothy, Hayes David L, Holmes David R, Asirvatham Samuel J

Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: Increasingly, autoidentification technology is being utilized in healthcare settings to make delivery of care cheaper, more efficient, and safer. However, the risk of electromagnetic interference (EMI) when using these autoidentification systems is not entirely clear. Thus, we sough to evaluate the likelihood of electromagnetic interference (EMI) in high technology healthcare areas with a low frequency, magnetic-field based wireless autoidentification protocol.

METHODS

Thirty-two devices were tested for EMI with an autoidentification technology whose maximal magnetic field output ranged from 8 to 800mgauss between November 2008 and March 2009. Testing was performed at Mayo Clinic (Rochester, MN) and Holy Cross Hospital (Miami, FL). All tests were started from 4 feet away and devices were approached in 1 foot increments with the signal antenna. EMI was characterized as light, significant, hazardous, or none per established testing standards. Significant to hazardous EMI was considered to potentially cause significant problems with clinical management of patients.

RESULTS

Of 32 devices, light to hazardous EMI was seen in 8 (25%). Hazardous EMI was seen in 2 12-lead ECG machines, significant EMI in fluoroscopy, echocardiograms, and 1 of 5 cardiac monitors, and light EMI in a defibrillator and cardiac monitor. Average magnetic field strengths experienced by the devices were similar at each distance regardless of the presence of EMI. No EMI was recorded with any device at distances greater than 4 feet, with no loss of signal fidelity at distances up to 17 feet.

CONCLUSION

A low frequency, magnetic field based wireless autoidentification technology induced distance dependent EMI in healthcare settings, making implementation potentially safe at antenna distances greater than 4 feet from clinical equipment. However, implementation requires rigorous in vivo testing to ensure the antenna is located a safe distance away.

Для цитирования: Kapa Suraj, Pierce Timothy, Hayes David L, Holmes David R, Asirvatham Samuel J. Electromagnetic interference of magnetic field based auto identification technologies in healthcare settings.. International journal of medical informatics. 2011. DOI: 10.1016/j.ijmedinf.2011.01.001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.