DOI 10.1067/mtc.2000.110189

Медицинская библиотека · The Journal of thoracic and cardiovascular surgery · 2000

Histologic and physiologic evaluation of skeletonized internal thoracic artery harvesting with an ultrasonic scalpel.

Higami T, Maruo A, Yamashita T, Shida T, Ogawa K

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

The safety and reliability of a method of skeletonized internal thoracic artery harvesting with an ultrasonic scalpel (Harmonic Scalpel; Ethicon Endo-Surgery, CVG, Cincinnati, Ohio) were evaluated.

METHODS

The mural branches of the internal thoracic artery were cut by means of 3 methods, differentiated by distance from the site of application of the Harmonic Scalpel blade to the internal thoracic artery. A total of 15 branches were cut from the internal thoracic artery at (0 mm) the origin (group I) or at 1 mm (group II) or 2 mm (group III) distal to the origin. Tissue preparations were examined for successful vessel closure and severity of tissue damage. The length of stump (L) and the length of tissue damage from the stump (D) were determined by a computer image analysis system, and pressure testing was performed to evaluate the physical strength of vessel closure.

RESULTS

In group I, 8 of the 15 branches exhibited discontinuity of the vascular wall structure, probably because of insufficient sealing of the divided section, and 12 of the 15 branches exhibited tissue denaturation on the internal thoracic artery wall adjacent to areas of origin, which was probably caused by the heat transferred from the branches during the process of coagulation. In groups II and III, continuity of wall structure of stumps suggestive of stable closure of branches was confirmed. The lengths of tissue damage from the stump (D) were 0.96, 0.58, and 0.63 mm in groups I, II, and III, respectively, and the lengths of intact area (L - D) in the corresponding groups were -0.78, 0.61, and 1.51 mm. The negative figure in group I indicates the presence of tissue damage in the internal thoracic artery itself. By contrast, in groups II and III the internal thoracic arteries were intact, with a safety margin of greater than 0.5 mm. On physiologic evaluation of vessel closure, 2 of the 24 (8.3%) branches burst under a pressure lower than 350 mm Hg because of insufficient vessel coagulation, but the remaining 22 branches (91.7%) remained intact under pressures up to 350 mm Hg.

CONCLUSION

The internal thoracic artery skeletonization method with an ultrasonic scalpel (Harmonic Scalpel: output level 2) appears to be a safe and reliable method of skeletonized internal thoracic artery harvesting when branches are sectioned at least 1 mm distal to their origin at a sufficiently slow speed.

Для цитирования: Higami T, Maruo A, Yamashita T, Shida T, Ogawa K. Histologic and physiologic evaluation of skeletonized internal thoracic artery harvesting with an ultrasonic scalpel.. The Journal of thoracic and cardiovascular surgery. 2000. DOI: 10.1067/mtc.2000.110189.