3-D computational analysis of the stress distribution on the leaflets after edge-to-edge repair of mitral regurgitation.

Сравнительное исследованиеОценочное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Votta Emiliano, Maisano Francesco, Soncini Monica, Redaelli Alberto, Montevecchi Franco M, Alfieri Ottavio

Аннотация

BACKGROUND AND AIM OF THE STUDY: Edge-to-edge repair is an effective, recently introduced method to correct mitral insufficiency by suturing the leaflets at the site of regurgitation, though durability of the method has not been proven. To overcome the limitations of the clinical approach, simulations may be used to predict clinical outcome. In this study, the mechanical stress acting on leaflets imposed by the edge-to-edge suture was evaluated as a means of assessing the clinical risk of late fibrosis or tissue degeneration.

METHODS: A 3-D finite element simulated the stress pattern following edge-to-edge repair. Valve behavior was evaluated both in systole and in diastole. Both 4-mm and 8-mm edge-to-edge sutures were simulated, as well as annular dilation.

RESULTS: Systolic simulations validated the model by comparison with previous models of the mitral valve. Diastolic stresses were negligible in the native mitral valve; after edge-to-edge repair (8-mm suture), circumferential and longitudinal stress values were 308 kPa and 489 kPa, respectively, and comparable with those observed at systolic peak (449 kPa and 617 kPa, respectively). With a 4-mm suture, longitudinal stresses decreased both close to the suture (-41.5%) and in the annular region (-68%), while circumferential stresses increased (+37%) close to the suture and decreased (-27%) in the annular region. A 20% dilation of the annulus was followed by increased stresses in the annular region and close to the suture.

CONCLUSION: Leaflet distortion and altered stress distribution occur on the leaflets after edge-to-edge repair. Diastolic peak stress values were comparable with those calculated in systole. The clinical implication is a doubled exposure of valve components to systolic stresses, as if the heart rate were doubled. The use of a prosthetic annuloplasty ring is favorable in the presence of annular dilation to reduce stresses acting on the leaflets after edge-to-edge repair.