DOI 10.1590/s0066-782x1998001200002
[Quantification of mitral prosthesis failure by different methods of evaluation by transesophageal echocardiography].
Аннотация
PURPOSE: To evaluate the different methods for grading mitral regurgitation (MR) by transesophageal echocardiography (TEE) in patients with clinical suspicion of mitral prosthesis dysfunction.
METHODS
Cardiac catheterization (Cath) was performed in 15 patients for grading the severity of prosthetic MR, divided in two groups based on the presence or absence of severe MR. Prosthetic MR was quantified by TEE using methods commonly used for MR of native valves: subjective assessment by color Doppler, objective assessment based on absolute jet area and on its relative area (jet area/left atrial area) and assessment based on the presence of systolic flow reversal in pulmonary vein.
RESULTS
Prosthetic MR was mostly transprosthetic (14 patients) and eccentric (11 patients). There was significant correlation (p < 0.05) between Cath and TEE for identification of severe MR based on subjective assessment and on the presence of systolic flow reversal in pulmonary vein. Identification based on absolute (jet area > 7 cm2) and relative (jet area > 35% of left atrial area) jet areas did not reveal significant correlation with the angiographic grade and showed clear underestimation by TEE when the last method was used. However, there was good correlation (p < 0.05) if relative jet areas > 30% were considered as cut point.
CONCLUSION
TEE correctly identified angiographic severe mitral prosthesis regurgitation, mainly by the presence of systolic flow reversal in pulmonary vein and subjective assessment. The estimation of severity of the prosthetic MR by absolute or relative jet area seems to be limited and should be used with caution due to eccentricity of the regurgitant jet, frequently seen in mitral prosthesis dysfunction.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.