DOI 10.1097/pcc.0000000000000947
Early Noninvasive Neurally Adjusted Ventilatory Assist Versus Noninvasive Flow-Triggered Pressure Support Ventilation in Pediatric Acute Respiratory Failure: A Physiologic Randomized Controlled Trial.
Аннотация
OBJECTIVE
Neurally adjusted ventilatory assist has been shown to improve patient-ventilator interaction in children with acute respiratory failure. Objective of this study was to compare the effect of noninvasive neurally adjusted ventilatory assist versus noninvasive flow-triggered pressure support on patient-ventilator interaction in children with acute respiratory failure, when delivered as a first-line respiratory support.
DESIGN
Prospective randomized crossover physiologic study.
SETTING
Pediatric six-bed third-level PICU.
PATIENTS: Eighteen children with acute respiratory failure needing noninvasive ventilation were enrolled at PICU admission.
INTERVENTIONS: Enrolled children were allocated to receive two 60-minutes noninvasive flow-triggered pressure support and noninvasive neurally adjusted ventilatory assist trials in a crossover randomized sequence.
MEASUREMENTS AND MAIN
RESULTS
Primary endpoint was the asynchrony index. Parameters describing patient-ventilator interaction and gas exchange were also considered as secondary endpoints. Noninvasive neurally adjusted ventilatory assist compared to noninvasive flow-triggered pressure support: 1) reduced asynchrony index (p = 0.001) and the number of asynchronies per minute for each type of asynchrony; 2) it increased the neuroventilatory efficiency index (p = 0.001), suggesting better neuroventilatory coupling; 3) reduced inspiratory and expiratory delay times (p = 0.001) as well as lower peak and mean airway pressure (p = 0.006 and p = 0.038, respectively); 4) lowered oxygenation index (p = 0.043). No adverse event was reported.
CONCLUSIONS
In children with mild early acute respiratory failure, noninvasive neurally adjusted ventilatory assist was feasible and safe. Noninvasive neurally adjusted ventilatory assist compared to noninvasive flow-triggered pressure support improved patient-ventilator interaction.