Depth-Dependent Corneal Biomechanical Properties in Normal and Keratoconic Subjects by Optical Coherence Elastography.
СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
De Stefano Vinicius S, Ford Matthew R, Seven Ibrahim, Dupps William J
Аннотация
PURPOSE: Compare depth-resolved biomechanical properties in normal and keratoconic corneas in live human subjects using optical coherence elastography (OCE).
METHODS: In a prospective series of normal and keratoconus (KC) eyes, a corneal perturbation was applied by a custom swept-source OCE system using a transparent flat lens coupled to force transducers. Cross-correlation was applied to track frame-by-frame OCT speckle displacement. Regional displacements for the anterior and posterior stroma were plotted in force versus displacement (k) graphs. A spatial biomechanical property ratio (ka/kp ) was defined by dividing the maximum total displacement by the maximum force for the anterior (ka ) and posterior cornea (kp) and was compared between normal and KC groups with the Mann-Whitney U test. Area under the receiver operating characteristics curve (AUROC) for differentiating normal and KC eyes was calculated for ka/kp , kmax, and thinnest point of corneal thickness (TPCT).
RESULTS: Thirty-six eyes were analyzed (21 eyes of 12 normal subjects and 15 KC eyes of 12 subjects). The ka/kp for the normal group was 1.135 ± 0.07 (mean ± standard deviation) and 1.02 ± 0.08 for the KC group (P < 0.001), indicating a relative deficit in anterior stromal stiffness in KC eyes. AUROC was 0.91 for ka /kp , 0.95 for kmax, and 1 for TPCT.
CONCLUSIONS: Significant differences in depth-dependent corneal biomechanical properties were observed between normal and KC subjects.
TRANSLATIONAL RELEVANCE: OCE was applied for the first time to human KC subjects and revealed alterations in the normal anterior-to-posterior stromal stiffness gradient, a novel and clinically accessible disease biomarker.
METHODS: In a prospective series of normal and keratoconus (KC) eyes, a corneal perturbation was applied by a custom swept-source OCE system using a transparent flat lens coupled to force transducers. Cross-correlation was applied to track frame-by-frame OCT speckle displacement. Regional displacements for the anterior and posterior stroma were plotted in force versus displacement (k) graphs. A spatial biomechanical property ratio (ka/kp ) was defined by dividing the maximum total displacement by the maximum force for the anterior (ka ) and posterior cornea (kp) and was compared between normal and KC groups with the Mann-Whitney U test. Area under the receiver operating characteristics curve (AUROC) for differentiating normal and KC eyes was calculated for ka/kp , kmax, and thinnest point of corneal thickness (TPCT).
RESULTS: Thirty-six eyes were analyzed (21 eyes of 12 normal subjects and 15 KC eyes of 12 subjects). The ka/kp for the normal group was 1.135 ± 0.07 (mean ± standard deviation) and 1.02 ± 0.08 for the KC group (P < 0.001), indicating a relative deficit in anterior stromal stiffness in KC eyes. AUROC was 0.91 for ka /kp , 0.95 for kmax, and 1 for TPCT.
CONCLUSIONS: Significant differences in depth-dependent corneal biomechanical properties were observed between normal and KC subjects.
TRANSLATIONAL RELEVANCE: OCE was applied for the first time to human KC subjects and revealed alterations in the normal anterior-to-posterior stromal stiffness gradient, a novel and clinically accessible disease biomarker.