Comparative Evaluation of Central Corneal Thickness in Cross-Linked Keratoconic Eyes.
Аннотация
PURPOSE: To compare the central corneal thickness (CCT) measurements measured by using the Pentacam Scheimpflug system with the CCT measured using ultrasound pachymetry (UP) and to compare the compatibility between the methods in normal, keratoconic, and cross-linked keratoconic corneas.
METHODS
The study included 50 eyes of 50 patients with keratoconus (keratoconus group), 50 eyes of 50 patients with progressive keratoconus who underwent corneal cross-linking treatment (CCL group), and 50 eyes of 50 healthy subjects (control group). Patients in the keratoconus and CCL groups were further classified into mild (mean keratometry [Km] ≤ 47 D) and moderate keratoconus subgroups (47.0 < Km < 52.0 D). CCT values were noted from the Pentacam Scheimpflug and UP.
RESULTS
The difference between the Pentacam and UP values was largest in the CCL group (-20.9 0.21.5 μm), followed by the keratoconus and control groups (-10.6 0.20.3 and 0.4 0.6.8 μm). The Pentacam and UP measurements were not comparable in the keratoconus and CCL groups (P = 0.001 and P < 0.001), whereas they were comparable in the control group (P = 0.62). In subgroup analysis, the 2 methods were comparable in the mild keratoconus subgroup (P = 0.12) and not comparable in the moderate keratoconus subgroup and in both mild and moderate subgroups of the CCL group (P = 0.001, P < 0.001 and P < 0.001).
CONCLUSIONS
Pentacam Scheimpflug tomography and UP can be used interchangeably in normal and mild keratoconic eyes, but not in moderate keratoconic and cross-linked keratoconic eyes. Pachymetry measurements from Scheimpflug must be interpreted with extreme caution with different methods before planning an invasive procedure in these eyes.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.