Evaluation of Biomechanically Corrected Intraocular Pressure Measurements in Keratoconus and Forme Fruste Keratoconus.
Аннотация
INTRODUCTION
Although biomechanically corrected intraocular pressure (bIOP) is available, the effectiveness of intraocular pressure (IOP) correction in keratoconus and forme fruste keratoconus (FFK) eyes has not been investigated.
OBJECTIVE
Evaluation of bIOP measurements in eyes with keratoconus and FFK.
METHODS
Forty-two eyes in 21 patients with keratoconus in one eye and FFK in the fellow eye were examined (KC/FFK group; mean age 24.62 ± 8.6 years; 16 males and 5 females). The control group consisted of 62 eyes in 31 unaffected subjects (mean age 26.26 ± 3.64 years; 15 males and 16 females). The bIOP was determined using a Scheimpflug-based tonometer (Corvis Scheimpflug Technology [Corvis ST®]) after measuring the IOP with a conventional non-contact tonometer (NIOP). The agreement between NIOP and bIOP values was examined using the Bland-Altman plot. The difference between NIOP and bIOP (bIOP correction amount) was compared between keratoconus and FFK eyes.
RESULTS
In the control group, there were no significant differences between right and left eyes in both NIOP and bIOP values (p = 0.975 and p = 0.224, respectively). In the KC/FFK group, NIOP values were significantly lower in the keratoconus eyes (9.93 ± 1.96 mm Hg) than in the FFK eyes (12.23 ± 3.03 mm Hg; p = 0.0003). There was no significant difference in bIOP values between the right and left eyes of the KC/FFK group (p = 0.168). The bIOP correction amount was significantly increased in keratoconus eyes (3.58 ± 2.12 mm Hg) compared to in FFK eyes (1.80 ± 3.32 mm Hg; p = 0.011).
CONCLUSIONS
For eyes with keratoconus and FFK, the bIOP method is effective to adjust IOP measurements based on corneal biomechanical properties.