Utility of a novel pacing guidewire in pre-implantation testing at different left ventricular sites in cardiac resynchronization therapy procedures.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Chan Ngai Yin, Choy Chi Chung, Lau Chun Leung, Lo Ying Keung, Chu Pui Shan, Yuen Ho Chuen, Lau Suet Ting

Аннотация

PURPOSE: This study aimed to evaluate the utility of a novel pacing guidewire in pre-implantation testing of different left ventricular (LV) sites during cardiac resynchronization therapy (CRT) procedures.

METHODS: Ten consecutive patients (8 male, mean age 65.8 ± 4.9) undergoing CRT procedures were studied. Pacing threshold and R-wave sensing measured by the guidewire and LV lead at different LV sites were compared.

RESULTS: Thirty sites (6 apical, 13 middle, and 11 basal; 15 lateral and 15 anterior) were tested. There was significant correlation between pacing threshold (r = 0.878, p < 0.0001), and R-wave sensing (r = 0.896, p < 0.0001) obtained by guidewire and those obtained by LV lead. Separating into lateral and anterior sites, significant correlation was also found in pacing threshold (lateral r = 0.658, p = 0.008; anterior r = 0.886, p < 0.0001) and R-wave sensing (lateral r = 0.887, p < 0.0001; anterior 0.865, p < 0.0001). For basal and middle sites, significant correlation was found in pacing threshold (basal r = 0.890, p < 0.0001; middle r = 0.878, p < 0.0001), and R-wave sensing (basal r = 0.930, p < 0.0001; middle r = 0.823, p < 0.001). No and borderline correlation was found in pacing threshold (r = 0.548, p = 0.26) and R-wave sensing (r = 0.835, p = 0.039), respectively, for apical sites. Concordance rate for the presence of phrenic nerve stimulation at high pacing output was 87%.

CONCLUSION: The accuracy of the novel pacing guidewire in pre-implantation testing in CRT procedures is site-dependent. There was good correlation with LV lead in the measurement of pacing threshold and R-wave sensing at basal and middle sites, but not apical sites. Presence of phrenic nerve stimulation can be predicted by guidewire testing with high accuracy.