Utility of a novel pacing guidewire in pre-implantation testing at different left ventricular sites in cardiac resynchronization therapy procedures.

Chan Ngai Yin, Choy Chi Chung, Lau Chun Leung, Lo Ying Keung, Chu Pui Shan, Yuen Ho Chuen, Lau Suet Ting

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: This study aimed to evaluate the utility of a novel pacing guidewire in pre-implantation testing of different left ventricular (LV) sites during cardiac resynchronization therapy (CRT) procedures.

METHODS

Ten consecutive patients (8 male, mean age 65.8 ± 4.9) undergoing CRT procedures were studied. Pacing threshold and R-wave sensing measured by the guidewire and LV lead at different LV sites were compared.

RESULTS

Thirty sites (6 apical, 13 middle, and 11 basal; 15 lateral and 15 anterior) were tested. There was significant correlation between pacing threshold (r = 0.878, p < 0.0001), and R-wave sensing (r = 0.896, p < 0.0001) obtained by guidewire and those obtained by LV lead. Separating into lateral and anterior sites, significant correlation was also found in pacing threshold (lateral r = 0.658, p = 0.008; anterior r = 0.886, p < 0.0001) and R-wave sensing (lateral r = 0.887, p < 0.0001; anterior 0.865, p < 0.0001). For basal and middle sites, significant correlation was found in pacing threshold (basal r = 0.890, p < 0.0001; middle r = 0.878, p < 0.0001), and R-wave sensing (basal r = 0.930, p < 0.0001; middle r = 0.823, p < 0.001). No and borderline correlation was found in pacing threshold (r = 0.548, p = 0.26) and R-wave sensing (r = 0.835, p = 0.039), respectively, for apical sites. Concordance rate for the presence of phrenic nerve stimulation at high pacing output was 87%.

CONCLUSION

The accuracy of the novel pacing guidewire in pre-implantation testing in CRT procedures is site-dependent. There was good correlation with LV lead in the measurement of pacing threshold and R-wave sensing at basal and middle sites, but not apical sites. Presence of phrenic nerve stimulation can be predicted by guidewire testing with high accuracy.

Для цитирования: Chan Ngai Yin, Choy Chi Chung, Lau Chun Leung, Lo Ying Keung, Chu Pui Shan, Yuen Ho Chuen, Lau Suet Ting. Utility of a novel pacing guidewire in pre-implantation testing at different left ventricular sites in cardiac resynchronization therapy procedures.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2011. DOI: 10.1007/s10840-011-9602-0.