Iterative reconstruction technique vs filter back projection: utility for quantitative bronchial assessment on low-dose thin-section MDCT in patients with/without chronic obstructive pulmonary disease.

Koyama Hisanobu, Ohno Yoshiharu, Nishio Mizuho, Matsumoto Sumiaki, Sugihara Naoki, Yoshikawa Takeshi, Seki Shinichiro, Sugimura Kazuro

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

The aim of this study was to evaluate the utility of the iterative reconstruction (IR) technique for quantitative bronchial assessment during low-dose computed tomography (CT) as a substitute for standard-dose CT in patients with/without chronic obstructive pulmonary disease.

METHODS

Fifty patients (mean age, 69.2; mean % predicted FEV1, 79.4) underwent standard-dose CT (150mAs) and low-dose CT (25mAs). Except for tube current, the imaging parameters were identical for both protocols. Standard-dose CT was reconstructed using filtered back-projection (FBP), and low-dose CT was reconstructed using IR and FBP. For quantitative bronchial assessment, the wall area percentage (WA%) of the sub-segmental bronchi and the airway luminal volume percentage (LV%) from the main bronchus to the peripheral bronchi were acquired in each dataset. The correlation and agreement of WA% and LV% between standard-dose CT and both low-dose CTs were statistically evaluated.

RESULTS

WA% and LV% between standard-dose CT and both low-dose CTs were significant correlated (r > 0.77, p < 0.00001); however, only the LV% agreement between SD-CT and low-dose CT reconstructed with IR was moderate (concordance correlation coefficient = 0.93); the other agreement was poor (concordance correlation coefficient <0.90).

CONCLUSIONS

Quantitative bronchial assessment via low-dose CT has potential as a substitute for standard-dose CT by using IR and airway luminal volumetry techniques.

KEY POINTS: • Quantitative bronchial assessment of COPD using low-dose CT is possible. • Airway luminal volumetry with iterative reconstruction is insusceptible to dose reduction. • Filtered back-projection is susceptible to the effect of dose reduction. • Wall area percentage assessment is easily influenced by dose reduction.

Для цитирования: Koyama Hisanobu, Ohno Yoshiharu, Nishio Mizuho, Matsumoto Sumiaki, Sugihara Naoki, Yoshikawa Takeshi, Seki Shinichiro, Sugimura Kazuro. Iterative reconstruction technique vs filter back projection: utility for quantitative bronchial assessment on low-dose thin-section MDCT in patients with/without chronic obstructive pulmonary disease.. European radiology. 2014. DOI: 10.1007/s00330-014-3207-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.