Predictive factors of difficult implantation procedure in cardiac resynchronization therapy.

Bisch Laurence, Da Costa Antoine, Dauphinot Virginie, Romeyer-Bouchard Cécile, Khris Lila, M'baye Alassane, Isaaz Karl

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: The usefulness of cardiac resynchronization therapy (CRT) in patients with congestive heart failure is offset by its long, user-dependent, and technical procedure. No studies have been published regarding factors related to CRT implantation procedure duration and X-ray exposure. Additionally, only a few studies have investigated the predictive factors of primary left ventricular (LV) lead implant failure. The aim of this prospective study was two-fold: (i) to evaluate the prevalence and predictive factors of prolonged CRT implantation procedure and (ii) to identify the predictive factors of primary LV lead implantation failure.

METHODS AND

RESULTS

Between November 2008 and September 2009, 128 consecutive patients underwent CRT implantation; of these, 22 patients (17.2%) were excluded because of CRT generator replacement. Population characteristics were a mean age of 69 +/- 10 years, 28.3% female, New York Heart Association class 3.2 +/- 0.3, LV ejection fraction (LVEF; 29 +/- 6%), and QRS width 146 +/- 23 ms. Cardiac resynchronization therapy implantation was attempted in 106 patients, and first LV lead implantation was obtained in 96 of 106 patients (90.5% primary success). Ten primary implantations failed (9.5%), due to unsuccessful LV lead implants. A second procedure was successfully attempted in six patients with a second more experienced operator (5.7%). Among the remaining four patients, one patient required a surgical epicardial LV lead implantation, and the implantation was not reattempted in the other three patients. The overall success rate of CRT system implantation was 96.2% (102 of 106 patients). Procedure parameters were as follows: LV threshold (1.4 +/- 0.9 V); LV wave amplitude (15 +/- 8 mV); LV impedance (874 +/- 215 ohm); median procedure time (skin to skin), 55 min (45-80); and median of procedure fluoroscopy time, 11 min (6.2-29). In 24 patients (22.6%), difficult procedures requiring >or=85 min of implantation duration occurred. By univariate analysis, predictive factors of difficult implantation were LV ejection fraction (25.6 +/- 6 vs. 30.2 +/- 8%; P = 0.02), LV end-diastolic diameter (72.4 +/- 11 vs. 66 +/- 11 mm; P = 0.01), LV end-systolic diameter (LVESD; 62 +/- 12 vs. 56 +/- 12 mm, P = 0.04), and the operator's experience (very experienced operator vs. less experienced operator, P = 0.006). By multivariate analysis, only primary LV lead implantation failure, LVESD, and operator's experience were independently associated with difficult procedures. In this patient subset with primary LV lead implant failure (n = 10), the only independent predictive factor was the LV end-systolic volume (P = 0.03).

CONCLUSION

In this study, the rate of difficult CRT device implantation procedures approached 25%. Both the degree of LV dysfunction and the operator's experience were independent predictors of surgical difficulties. Left ventricular end-systolic volume was the only independent predictor of primary LV lead implant failure.

Для цитирования: Bisch Laurence, Da Costa Antoine, Dauphinot Virginie, Romeyer-Bouchard Cécile, Khris Lila, M'baye Alassane, Isaaz Karl. Predictive factors of difficult implantation procedure in cardiac resynchronization therapy.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2010. DOI: 10.1093/europace/euq146.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.