DOI 10.1109/iembs.2009.5333095

Медицинская библиотека · Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2009

Simulation of AV hysteresis pacing using an integrated dual chamber heart and pacer model.

Lian Jie, Garner Garth, Kratschmer Hannes, Mussig Dirk

СтатьяПеревод пока не подготовлен

Аннотация

Long term right ventricular apical pacing has been known to have adverse effects in cardiac function. The AV hysteresis (AVH) is a feature existing in many dual-chamber cardiac pacemakers that aims to minimize the right ventricular pacing, but its clinical efficacy remains inconclusive due to conflicting evidence from different studies. We have recently developed a novel integrated dual-chamber heart and pacer (IDHP) model, which can simulate various interactions between intrinsic heart activity and extrinsic cardiac pacing. In this study, we use the IDHP model to simulate various atrio-ventricular (AV) conduction pathologies, and to investigate the effects of an AVH algorithm on reducing right ventricular pacing. Our results show that the efficacy of AVH is dependent on the underlying cardiac conditions. While it can preserve intrinsic conduction during minor or moderate first degree AV block, its efficacy is reduced at higher degree AV block conditions. This pilot study further supports using the IDHP model to design and evaluate more advanced pacemaker algorithms for therapeutic interventions.

Для цитирования: Lian Jie, Garner Garth, Kratschmer Hannes, Mussig Dirk. Simulation of AV hysteresis pacing using an integrated dual chamber heart and pacer model.. Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference. 2009. DOI: 10.1109/iembs.2009.5333095.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.