[Phrenic nerve stimulation in biventricular cardiac pacemakers].
Аннотация
BACKGROUND
Phrenic nerve stimulation (PNS) complicates the positioning of the left ventricle lead. We present a step-by-step approach to correct PNS during implantation, as established in our daily routine.
METHODS
The incidence of PNS, its successful correction, and long-term results (median 27 months) were analyzed retrospectively in 266 lead positions.
RESULTS
Phrenic nerve stimulation occurred in 13.9% of the lead positions. Multivariate analysis (P<0.02) showed that PNS only depended on the place of stimulation (coronary sinus side branch). Lead type, CRT indication, and patient's sex had no significant correlation. Following the step-by-step approach presented here, PNS was corrected satisfactorily in all cases.
CONCLUSIONS
Approach in case of PNS: 1. push or pull the lead within the same vein, 2. change to a different vein, 3. maintain position in case of a safe distance between the phrenic nerve and the pacing threshold, 4. change the lead type to achieve stable anchorage at adequate positions, 5. use a device featuring electronic repositioning.