Entrance skin dose during radiofrequency catheter ablation for tachyarrhythmia: a multicenter study.

Kidouchi Takashi, Suzuki Shigeru, Furui Shigeru, Mitani Haruo, Nitta Junichi, Matsumoto Kunihiro, Ohtomo Kuni

Сравнительное исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

To assess the entrance skin dose (ESD) during radiofrequency catheter ablation procedures for tachyarrhythmia including atrial fibrillation (Af).

METHODS

This study focused on 99 consecutive patients who underwent procedures for tachyarrhythmia (Af; n = 34, non-Af; n = 65) in three institutions. The non-Af group included atrial flutter, atrial tachycardia, paroxysmal supraventricular tachycardia, ventricular tachycardia, ventricular premature contraction, atrial premature contraction, atrioventricular nodal reentry tachycardia, and Wolff-Parkinson-White syndrome. In two of the three institutions, the procedures were performed for both Af and non-Af. The ESDs were measured using 100 radiosensitive indicators attached to the back of each patient's jacket at 5-cm intervals. For statistical analyses, multiple regression analysis (the dependent variable, Max-ESD; and the independent variables, dose area product [DAP], total fluoroscopic time [TFT], body mass index, etc.), Pearson's correlation test, and the Mann-Whitney test were employed.

RESULTS

The overall averages for the TFTs, the DAPs, and the Max-ESDs were 49.9 ± 28.2 minutes, 71.2 ± 73.7 Gy cm(2) , and 0.57 ± 0.51 Gy, respectively. DAP was positively related to the Max-ESD and was significant in stepwise multiple regression analysis (P < 0.0001). There was a significant association between TFT and Max-ESD in five of the six kinds of angiographic unit, and between DAP and Max-ESD in all three systems with available DAP measures. In one institution, TFT, DAP, and Max-ESD differed significantly between the Af and non-Af groups (P = 0.0002, P < 0.0001, and P < 0.0001).

CONCLUSIONS

During the cardiac catheter ablation, ESDs of only a few patients exceeded the thresholds of radiation skin injuries, and the DAP proved useful to estimate each patient's Max-ESD.

Для цитирования: Kidouchi Takashi, Suzuki Shigeru, Furui Shigeru, Mitani Haruo, Nitta Junichi, Matsumoto Kunihiro, Ohtomo Kuni. Entrance skin dose during radiofrequency catheter ablation for tachyarrhythmia: a multicenter study.. Pacing and clinical electrophysiology : PACE. 2011. DOI: 10.1111/j.1540-8159.2010.03016.x.