Status of Oxygen Monitoring in Four Selected Special Care Newborn Units in India.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Sabherwal Shalinder, Gilbert Clare, Foster Allen, Kumar Praveen
Аннотация
OBJECTIVES: To investigate the status of oxygen monitoring in Special Newborn Care Units.
METHODS: Observations were made and records reviewed of infants on oxygen in all four Special Newborn Care Units of a state delivering a model program for retinopathy of prematurity. Multiple choice questions were administered to nurses, semi-structured interviews conducted with pediatricians, ophthalmologists and senior nurses.
RESULTS: All units had more than 100% occupancy. The number of functioning pulse oximeters was 73% of that recommended. None of the units had air-oxygen blenders. The upper oxygen saturation alarm was set accurately only for 1 out of 18 babies receiving oxygen and none of the infants had continuous saturation monitoring. 84% of nurses did not know optimal oxygen saturation targets. Most interviewees attributed suboptimal care to overcrowding.
CONCLUSION: Compressed air, air-oxygen blenders, sufficient functioning pulse oximeters, rational admission policies and training of nurses are needed to improve oxygen related practices.
METHODS: Observations were made and records reviewed of infants on oxygen in all four Special Newborn Care Units of a state delivering a model program for retinopathy of prematurity. Multiple choice questions were administered to nurses, semi-structured interviews conducted with pediatricians, ophthalmologists and senior nurses.
RESULTS: All units had more than 100% occupancy. The number of functioning pulse oximeters was 73% of that recommended. None of the units had air-oxygen blenders. The upper oxygen saturation alarm was set accurately only for 1 out of 18 babies receiving oxygen and none of the infants had continuous saturation monitoring. 84% of nurses did not know optimal oxygen saturation targets. Most interviewees attributed suboptimal care to overcrowding.
CONCLUSION: Compressed air, air-oxygen blenders, sufficient functioning pulse oximeters, rational admission policies and training of nurses are needed to improve oxygen related practices.