Circulation. Arrhythmia and electrophysiology · 2020

Automated Noncontact Ultrasound Imaging and Ablation System for the Treatment of Atrial Fibrillation: Outcomes of the First-in-Human VALUE Trial.

СтатьяПоддержка исследования: вне правительства СШАКлиническое исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Turagam Mohit K, Petru Jan, Neuzil Petr, Kakita Ken, Kralovec Stepan, Harari David, Phillips Patrick, Piazza Danielo, Whang William, Dukkipati Srinivas R, Reddy Vivek Y

Аннотация

BACKGROUND: Catheter ablation for atrial fibrillation (AF) using point-by-point radiofrequency energy or single-application one-shot balloons is either technically challenging or have limited ability to accommodate variable patient anatomy to achieve acute and durable pulmonary vein (PV) isolation. A novel ablation system employs low intensity collimated ultrasound (LICU)-guided anatomic mapping and robotic ablation to isolate PVs. In this first-in-human, single-center, multioperator trial, VALUE trial (VytronUS Ablation System for Treatment of Paroxysmal Atrial Fibrillation; NCT03639597) in patients with paroxysmal atrial fibrillation, this LICU system was evaluated to determine its safety, effectiveness in PV isolation, and freedom from recurrent atrial arrhythmias.

METHODS: In the enrolled 52 patients with paroxysmal atrial fibrillation, ultrasound M-mode-based left atrial anatomies were successfully created, and ablation was performed under robotic control along an operator-defined lesion path. The LICU system software advanced over the course of the study: the last 13 patients were ablated with enhanced software.

RESULTS: Acute PV isolation was achieved in 98% of PVs-using LICU-only in 77.3% (153/198) of PVs and requiring touch-up with a standard radiofrequency ablation catheter in 22.7% (45/198) PVs. The touch-up rate decreased to 5.8% (3/52) in patients undergoing LICU-ablation with enhanced software. Freedom from atrial arrhythmia recurrence was 79.6% (39/49 patients) at 12 months or 92.3% (12/13 patients) with the enhanced software. Major adverse events occurred in 3 patients (5.8%): one had transient diaphragmatic paralysis, one vascular access complication, and one had transient ST-segment elevation from air-embolism, without sequelae.

CONCLUSIONS: In this first-in-human study, low- intensity collimated ultrasound-guided anatomic mapping and robotic ablation allows PV isolation with good chronic safety; PV isolation success is improving with device enhancements. Registration: URL: https://www.clinicaltrials.gov; Unique identifier: NCT03639597.