DOI 10.1016/j.ipej.2019.12.014
The efficacy of the LinoxSmart DX ICD lead from a single center experience.
Аннотация
PURPOSE: The Biotronik LinoxSmart DX implanted cardioverter defibrillator (ICD) lead is a novel VDD lead with the advantage of integrated atrial sensing dipole combined with a special augmentation and filtering mechanisms. We sought to determine the efficacy of the Biotronik LinoxSmart DX ICD lead.
METHODS
Non-randomized consecutive patients implanted with Biotronik LinoxSmart DX lead at Sheba Medical Center were included in this study. Electrical parameters and arrhythmic events were recorded during follow up of one year.
RESULTS
Seventy-three patients (69 males (94.5%), mean age 61 ± 12 years) were included. All patients were successfully implanted with a Biotronic VR-T DX device and LinoxSmart DX ICD lead (DX-17 in 37% and DX-15 in 63% patients). Mean P wave amplitude at time of implantation was 3.66 ± 2.9 mV and improved significantly throughout the follow-up (5.29 ± 4.39 mV, p = 0.009). Appropriate atrial sensing (defined as P wave amplitude of ≥0.8 mV) rate of 100% at implantation significantly decreased to 89% (p = 0.015) at 12 months. Three out of 67 (4.5%) patients without a known history of atrial fibrillation had documented new onset paroxysmal atrial fibrillation. Appropriate shocks occurred in 4 (5.5%) patients. One patient with atrial sensing less than 0.4 mV had inappropriate shock.
CONCLUSIONS
Among patients implanted with the Biotronik LinoxSmart DX ICD lead in our single center, appropriate atrial sensing rate decreased over 12 months. Larger studies are needed to evaluate the reliability of long term appropriate atrial sensing.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.