OBJECTIVE

The aim of this research was to investigate the relationship between atrial fibrillation (AF) recurrence and second generation cryoballoon ablation (CBA) procedural parameters in patients with non-valvular paroxysmal AF (PAF).

METHODS

A total of 131 patients with a PAF diagnosis who underwent second-generation CBA (59 male; mean age: 55.2±10.6 years) were enrolled. Recurrence was defined as the detection of AF on a 12-lead electrocardiography (ECG) recording, or an attack lasting at least 30 seconds observed on Holter ECG records. CBA procedural data and echocardiographic findings were recorded and compared.

RESULTS

After 1 year of follow-up, AF recurrence was detected in 27 patients. Patients with recurrence were older and had higher rates of hypertension and diabetes (p<0.05 for both). Left atrial diameter, left atrial volume (LaV), left atrial volume index, and the averaged warming angle (calculated by combining lowest temperature point and balloon temperature at 20°C point) were significantly higher in the recurrence group. Balloon warming time was significantly longer in the non-recurrence group (p<0.001). In binary logistic regression analysis, the averaged warming angle (odds ratio [OR]: 1.559, 95% confidence interval [CI]: 1.342-1.811; p<0.001) and LaV (OR: 1.063, 95% CI: 1.028-1.100; p<0.001) were found to be independent parameters for predicting recurrence. The cutoff value of the warming angle obtained with ROC curve analysis was 50° for the prediction of recurrence (sensitivity: 94.3%, specificity: 88.5%, area under the curve: 0.909; p<0.001). The cutoff value of LaV obtained by ROC curve analysis was 53.5 for prediction of recurrence (sensitivity: 77.8%, specificity: 74.5%; p<0.001).

CONCLUSION

Measurement of balloon warming angle during CBA and increased LaV may predict the AF recurrence.

Для цитирования: Demirtaş Abdullah Orhan, İçen Yahya Kemal, Dönmez Yurdaer, Koca Hasan, Kaypakli Onur, Şahin Durmuş Yıldıray, Koç Mevlüt. Effect of cryoballoon ablation parameters on recurrence in patients with paroxysmal atrial fibrillation.. Turk Kardiyoloji Dernegi arsivi : Turk Kardiyoloji Derneginin yayin organidir. 2019. DOI: 10.5543/tkda.2019.88469.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.